connecting cables and employ industry-standard ESD prevention measures during
installation, maintenance, and operation.
Figure 4. Installation of the NI 5668R Cables and 50
Ω
Loads
ACCESS
ACTIVE
4
3
1
5
6
2
3
1. SMA-to-SMA Cable, Labeled V
2. SMA-to-SMA Cable, Labeled U
3.
50 Ω
Load
4. SMA-to-SMA Cable, Labeled Y
5. SMA-to-SMA Cable, Labeled X
6. SMA-to-SMA Cable, Labeled W
1.
Using the three SMA-to-SMA coaxial cables labeled W, X, and Y, connect the LO1,
LO2, and LO3 connectors on the NI 5653 front panel to the associated LO1 IN, LO2 IN,
and LO3 IN connectors on the NI 5606 front panel.
2.
Using the SMA-to-SMA cable labeled U, connect the LO2 OUT connector on the
NI 5606 front panel to the CLK IN connector on the NI 5624R front panel.
3.
Using the SMA-to-SMA cable labeled V, connect the IF OUT connector on the NI 5606
front panel to the IF IN connector on the NI 5624R front panel.
4.
Install a 50 Ω load on each of the NI 5606 module LO1 OUT, LO3 OUT, and
IF IN / BIAS OUT connectors.