5.

Carefully complete tightening all SMA connectors to 1 N · m using an appropriate torque
wrench (not included) or torque screwdriver and SMA driver bit. Tighten only until the
wrench clicks.

Caution

 

Incorrect torque at SMA connections can degrade signal fidelity and

phase-locked loop (PLL) performance and may cause insertion loss. Use an
appropriate torque wrench or torque screwdriver and SMA driver bit to ensure
all SMA connections are properly torqued. SMA connectors for connections to
external equipment may require torque different from 1 N · m, depending on
the connector type, material, and manufacturer.

6.

Power on the PXI Express chassis and controller system.

7.

Verify that the ACCESS LED is on for all modules.

Related Information

Troubleshooting

 on page 23

If an issue persists after you complete a troubleshooting procedure, contact NI technical
support or visit 

ni.com/support

.

Refer to the NI RF Vector Signal Analyzers Help for more information about how to connect
the NI 5663E using an external clock, how to connect multiple NI 5663E modules, or how to
connect the NI 5663E to the NI PXIe-5673E RF vector signal generator.

Direct Connections to the NI 5601

The NI 5601 is a precision RF instrument that is sensitive to ESD and transients. Ensure you
are making proper direct connections to the NI 5601 to avoid damaging the device.

Caution

 

Apply external signals only while the NI 5601 is powered on. Applying

external signals while the device is powered off may cause damage.

Ensure you are properly grounded when manipulating cables or antennas connected to the
NI 5601 RF IN connector.

If you are using noninsulated devices, such as an RF antenna, ensure the devices are
maintained in a static-free environment.

If you are using an active device, such as a preamplifier or switch routed to the NI 5601
RF IN connector, ensure that the device cannot generate signal transients greater than the
RF and DC specifications of the NI 5601 RF IN connector.

10

  | 

ni.com

  | 

NI PXIe-5663E Getting Started Guide

Summary of Contents for PXIe-5663E

Page 1: ...5601 NI 5601 RF downconverter module NI PXIe 5622 NI 5622 IF digitizer module NI PXIe 5652 NI 5652 LO source module Caution The protection provided by this product may be impaired if it is used in a...

Page 2: ...adation install and use this product in strict accordance with the instructions in the product documentation Furthermore any changes or modifications to the product not expressly approved by National...

Page 3: ...age Caution Never touch the exposed pins of connectors Note Do not install a device if it appears damaged in any way 3 Unpack any other items and documentation from the kit Store the device in the ant...

Page 4: ...Rigid SMA to SMA Cable Labeled D Part Number 198776A 01 5 Medium Semi Rigid SMA to SMA Cable Labeled C Part Number 198775A 01 6 Flexible SMA to SMA Cable Part Number 190412B 03 7 SMA Driver Bit Part N...

Page 5: ...ing the NI 5663E in meets the following specifications Operating ambient temperature IEC 60068 2 1 IEC 60068 2 2 0 C to 55 C Operating relative humidity IEC 60068 2 56 10 to 90 noncondensing Maximum a...

Page 6: ...ls to test receivers NI RFSA SFP Controls analyzes and presents data similar to stand alone RF vector signal analyzers The NI RFSA SFP operates on the PC so it provides processing storage and display...

Page 7: ...chassis 3 Inspect the slot pins on the chassis backplane for any bends or damage prior to installation Do not install a module if the backplane is damaged 4 Remove the black plastic covers from all th...

Page 8: ...ility and also electrically connects the front panel to the chassis which can improve the signal quality and electromagnetic performance 11 Cover all empty slots using filler panels or fill using slot...

Page 9: ...er pin of a medium semi rigid SMA to SMA cable labeled C connect the RF OUT connector on the NI 5652 LO source front panel to the LO IN connector on the NI 5601 RF downconverter front panel 2 Using th...

Page 10: ...F Vector Signal Analyzers Help for more information about how to connect the NI 5663E using an external clock how to connect multiple NI 5663E modules or how to connect the NI 5663E to the NI PXIe 567...

Page 11: ...nnectors and two LEDs Figure 5 NI 5601 RF Downconverter Module Front Panel 0 dBm NOM RF IN IF OUT LO IN LO OUT 10 MHz 6 6 GHz ACCESS ACTIVE ALL PORTS 5 VDC MAX 50 0 dBm NOM 20 dBm MAX 20 dBm MAX 20 dB...

Page 12: ...analog RF input signal to be measured by the NI 5663 IF OUT Connects to the IF IN connector on the NI 5622 IF digitizer module front panel This connector is the output terminal for the frequency tran...

Page 13: ...in order to control the device GREEN The module is ready to be programmed by NI RFSA ACTIVE Indicates the module state OFF The module is in a state other than thermal shutdown or power supply failure...

Page 14: ...t panel contains four connectors and two LEDs Figure 6 NI 5622 IF Digitizer Module Front Panel ACCESS ACTIVE CLK IN CLK OUT PFI 1 ESD SENSITIVE TTL 6 3 Vp p MAX 2 Vp p NOM 20 dBm MAX 50 50 50 16 Bit I...

Page 15: ...Module Front Panel Connectors Connector Use IF IN Connects to the IF OUT connector on the NI 5601 RF downconverter module front panel This connector is the input terminal for a frequency translated IF...

Page 16: ...to be programmed by NI RFSA ACTIVE Indicates the module state OFF The module is in a quiescent state AMBER The module is waiting for the Advance Trigger from the configuration list GREEN The module is...

Page 17: ...0 5 W MAX Reverse Power RF Signal Generator ACCESS ACTIVE Table 8 Device Front Panel Icon Definitions Refer to the user documentation for required maintenance measures to ensure user safety and or pre...

Page 18: ...d a problem with a power rail AMBER The module is being accessed Accessed means that the device setup registers are being written to in order to control the device GREEN The module is ready to be prog...

Page 19: ...identified as not configured in the configuration tree b In the Associated Devices section select the appropriate module from each system component drop down listbox For the NI 5663E you must associat...

Page 20: ...library files located in one of the following directories Windows 10 32 bit 8 1 32 bit 7 32 bit NIDir Shared ExternalCompilerSupport C include Windows 10 64 bit 8 1 64 bit 7 64 bit 32 bit installation...

Page 21: ...s x86 IVI Foundation IVI Microsoft Visual C C Use the header files located in the IVIROOTDIR32 Include directory and import library files located in the IVIROOTDIR32 Lib directory where IVIROOTDIR32 i...

Page 22: ...an Application with Microsoft Visual C and C topic of the NI RF Vector Signal Analyzers Help if you prefer to manually add all required include and library files to the project Programming Examples Lo...

Page 23: ...ples Troubleshooting If an issue persists after you complete a troubleshooting procedure contact NI technical support or visit ni com support Installation Why Is the ACCESS LED Off When the Chassis Is...

Page 24: ...ditions appear in the list If you are using an MXI controller right click PCI to PCI Bridge and select Properties from the shortcut menu to verify that the bridge is enabled What Should I Do if the Mo...

Page 25: ...nal analyzer 2 Verify the MAX device association for your NI 5663E 3 If you have verified the interconnections and associations of your device and you are still unable to launch the NI RFSA SFP try un...

Page 26: ...i com services NI Community ni com community Located at ni com gettingstarted Located at ni com manuals Located using the NI Example Finder Tip The NI RF Vector Signal Analyzers Help is an HTML versio...

Page 27: ...te for your product at ni com calibration NI corporate headquarters is located at 11500 North Mopac Expressway Austin Texas 78759 3504 NI also has offices located around the world For telephone suppor...

Page 28: ...arty legal notices in the readme file for your NI product Refer to the Export Compliance Information at ni com legal export compliance for the NI global trade compliance policy and how to obtain relev...

Reviews: