NI PXIe-4480/4481 Calibration Procedure
|
© National Instruments
|
9
Charge Gain Accuracy Verification/Adjustment
(PXIe-4480 only)
Charge gain verification and adjustment is conducted one channel at a time. Figure 4 illustrates
the connections for charge gain accuracy verification/adjustment.
Figure 4.
Charge Gain Accuracy Connections
Note
Secure the cable to the module front panel using the cable connector
jackscrews.
1
Calibrator
2
Banana Plug Adapter
3
10-32 to BNC Interface Cable
4
Calibration Capacitor
5
BNC Cable
6
InfiniBand 12X-6BNC Cable
7
PXIe-4480
1
2
3
6
7
5
4