NI PXIe-4480/4481 Calibration Procedure
|
© National Instruments
|
7
Voltage Gain Accuracy Verification/Adjustment
Voltage gain verification and adjustment can be conducted one channel at a time, all six channels
simultaneously, as well as any number in between. Figure 2 illustrates the connections for
parallel channel operation, while Figure 3 illustrates the connections for sequential channel
operation.
Figure 2.
Voltage Gain Accuracy/Flatness Connections (Six Channels)
Note
Secure the cable to the module front panel using the cable connector
jackscrews.
1
Calibrator
2
Banana Plug Adapter
3
BNC Cable
4
BNC T-Connector
5
InfiniBand 12X-6BNC Cable
6
PXIe-4480/4481
1
3
2
6
5
4