PXIe-4468 Calibration Procedure
10
ni.com/services
Figure 1.
Initial Connection for AI Offset-BNC
1. Calibrator
3. BNC – BNC Cable
2. Banana – BNC Adapter
4. PXIe-4468 with BNC
Figure 2.
Initial Connection for AI Offset-mXLR
1. Calibrator
3. BNC – mXLR Cable
2. Banana – BNC Adapter
4. PXIe-4468 with mXLR
Verification Procedure
Complete the following procedure to verify the AI offset accuracy.
Repeat 2 times, once for each channel.
1.
Connect the AI channel under test to the Calibrator as shown in
or