PXIe-4468 Calibration Procedure
18
ni.com/services
Figure 4.
Initial Connection for IEPE Current-mXLR
1. DMM
3. BNC – mXLR Cable
2. Banana – BNC Adapter
4. PXIe-4468 with mXLR
Verification Procedure
Complete the following procedure to verify the IEPE current.
Repeat 2 times, once for each channel.
1.
Connect the AI channel under test to the DMM as shown in
Repeat 3 times, once for each IEPE current setting.
2.
Set the DMM to the AMPS function and the 10 mA range for the 4 mA
and 10 mA test points, and the 100 mA range for the 20 mA test point.
3.
Create and configure an AI Voltage task using the values in
Table 10:
IEPE Current Verification Configuration
Configuration
Value
Physical Channels
Dev_name
/ai
x
Gain*
-20 (dB)
Coupling
AC