National Instruments PXIe-4357 Calibration Procedure Download Page 5

© National Instruments Corporation

|

5

Refer to Table 2 for the analog signal names of the TB-4357.

Verification

Verification determines how well the module is meeting its specifications. By completing this 
procedure, you can see how the module has drifted over time, which helps you determine the 
appropriate calibration interval for your application. The NI PXIe-4357 has 20 independent 
analog input channels. You can conduct verification on any or all of them, depending upon your 
desired test coverage.

4-Wire Accuracy Verification

Complete the following steps to test the 4-wire accuracy of the module.
1.

Set the calibrator to standby mode (STBY).

2.

Make the following connections to the TB-4357 for the channel you want to verify. Refer 
to Figure for visual reference. Refer to the 

Connecting the Calibrator to the Module

 

section for the pin assignments and signal names of the TB-4357.

Figure 2.  

4-Wire Accuracy Verification Connections

3.

Zero the calibrator with the ohms-only zero (OHMS ZERO) operation. Refer to the 
calibrator documentation for more information about zeroing the calibrator.

4.

Set the calibrator output for 4-wire compensation (COMP 4-wire).

5.

Set the calibrator resistance output to a 0

Ω

.

6.

Set the calibrator to operate mode (OPR).

7.

Create a task.

Table 2.  

TB-4357 Analog Signal Names

Signal Names

Signal Descriptions

EX–

Excitation current return

EX+

Positive excitation current output

AI–

Negative analog input

AI+

Positive analog input

C

a

li

b

r

a

tor

HI

AUX HI

AUX LO

LO

TB-4

3

57

CH X

EX+

AI+
AI–

EX–

Summary of Contents for PXIe-4357

Page 1: ...cedure 7 Specifications 9 Where to Go for Support 10 Software Requirements Install NI DAQmx 9 5 or later on the calibration computer NI DAQmx includes high level function calls to simplify the task of...

Page 2: ...use the NI DAQmx driver The NI DAQmx C Reference Help includes information about the functions in the driver Calibration Interval National Instruments recommends a calibration interval of one year fo...

Page 3: ...mance of the NI PXIe 4357 The calibration process consists of the following steps 1 Initial Setup Install the module and configure it in Measurement Automation Explorer MAX 2 Verification Verify the e...

Page 4: ...on describing the required module connections Each channel consists of a group of four pins specific to that channel Before connecting or disconnecting the calibrator from the module always set the ca...

Page 5: ...the following connections to the TB 4357 for the channel you want to verify Refer to Figure 2 for visual reference Refer to the Connecting the Calibrator to the Module section for the pin assignments...

Page 6: ...cation 15 Compare the measured result for each value to the Upper Limit and Lower Limit values in Table 4 If the results are all between these values the module passes the test 16 Set the calibrator t...

Page 7: ...etermine the gain and offset adjustment calibration constants Complete the following steps to test the accuracy of the module 1 Set the calibrator to standby mode STBY 2 Make the following connections...

Page 8: ...ration VI The default password is NI 8 Call the DAQmx Adjust 4357 Calibration VI 9 Set the calibrator to output a resistance of 350 This is the second calibration point 10 Call the DAQmx Adjust 4357 C...

Page 9: ...module you must reverify the accuracy of the module To do this repeat the Verification section Specifications The specifications in Table 5 are provided for reference only refer to the NI PXIe 4357 S...

Page 10: ...Information at ni com legal export compliance for the National Instruments global trade compliance policy and how to obtain relevant HTS codes ECCNs and other import export data Where to Go for Suppo...

Reviews: