NI PXIe-4340 Calibration Procedure

|

© National Instruments

|

5

Each channel consists of two terminal connections specific to that channel as shown in Table 2. 
You can verify the accuracy for any or all channels depending on the desired test coverage.

Refer to Table 2 for the analog signal names of the TB-4340.

Test Conditions

The following setup and environmental conditions are required to ensure the PXIe-4340 meets 
published specifications:

Keep connections to the PXIe-4340 as short as possible. Long cables and wires act as 
antennas, picking up extra noise that can affect measurements.

Verify that all connections to the TB-4340 are secure.

Use shielded copper wire for all cable connections to the TB-4340. Use twisted-pair wire 
to eliminate noise and thermal offsets.

Maintain an ambient temperature of 23 °C ±5 °C. The PXIe-4340 temperature will be 
greater than the ambient temperature. Maintain an internal device temperature range of 

T

cal

± 5 °C for the PXI-4070.

1

Keep relative humidity below 80%.

For valid test limits, the device must be within ±5 °C and 24 hours of the last 
self-calibration.

Table 2.  

TB-4340 Analog Signal Names

Signal Name

Signal Description

Excitation Verifi

Positive excitation terminal for gross checks

Excitation Verification -

Negative excitation terminal for gross checks

EX+

Positive excitation terminal

EX-

Negative excitation terminal

RS+

Positive remote sense terminal

RS-

Negative remote sense terminal

AIGND

Analog input ground

AI+

Positive input voltage terminal

AI-

Negative input voltage terminal

CT

LVDT center tap connection

1   

T

cal

 is the internal device temperature recorded by the PXI-4070 at the completion of the last 

self-calibration.

Summary of Contents for PXIe-4340

Page 1: ...rldwide Support and Services 10 Software Calibrating the NI PXIe 4340 requires the installation of NI DAQmx on the calibration system Driver support for calibrating the PXIe 4340 was first available i...

Page 2: ...nformation NI PXIe 4340 Device Specifications PXIe 4340 specifications and calibration interval NI DAQmx Readme Operating system and application software support in NI DAQmx NI DAQmx Help Information...

Page 3: ...not available select a substitute using the requirements listed in Table 1 Table 1 Recommended Equipment for PXIe 4340 Verification Equipment Recommended Model Requirements DMM PXI 4070 Use a DMM that...

Page 4: ...he TB 4340 Figure 1 TB 4340 Circuit Board Parts Locator Diagram C1 B1 A1 COPYRIGHT 2015 S N 159026B 01L NI TB 4340 EX EX RS RS AI AI CT AIGND CHANNEL 0 EXCITATION VERIFICATION EX EX RS RS AI AI CT AIG...

Page 5: ...minate noise and thermal offsets Maintain an ambient temperature of 23 C 5 C The PXIe 4340 temperature will be greater than the ambient temperature Maintain an internal device temperature range of Tca...

Page 6: ...the device name as it appears in MAX Accuracy Verification The following performance verification procedures describe the sequence of operations and provides the test points required to verify the PXI...

Page 7: ...zero Once ADC Calibration On Input Impedance 10 M DC Noise Rejection High Order Number of Averages 10 Power Line Frequency Dependent upon the local power line characteristics Table 4 PXI 4070 Voltage...

Page 8: ...hese values the devices passes the DC Offset test at this Excitation Voltage Table 5 PXIe 4340 AI LVDT Measurement Setup Configuration Value Channel Name Dev1 aix where x refers to the channel number...

Page 9: ...8 Leave all other pins disconnected 2 Configure the channel as shown in Table 5 for 7 Vrms excitation 3 Perform the configured AI LVDT measurement and average the acquired measurements 4 Compare the r...

Page 10: ...ompleted initialize and close a calibration sessions for Dev1 using the DAQmx Initialize External Calibration VI and the DAQmx Close External Calibration VI Specifications Refer to the NI PXIe 4340 De...

Reviews: