background image

2

|

ni.com

|

PXIe-4310 Calibration Procedure

Documentation

Consult the following documents for information about the PXIe-4310, NI-DAQmx, and your 
application software. All documents are available on 

ni.com

, and help files install with the 

software.

PXIe-4310 and TB-4310 (10V)/TB-4310 (600V) Getting Started Guide and Terminal Block 
Specifications

NI-DAQmx driver software installation and hardware setup.

PXIe-4310 User Manual

PXIe-4310 usage and reference information.

PXIe-4310 Specifications

PXIe-4310 specifications and calibration interval.

NI-DAQmx Readme

Operating system and application software support in NI-DAQmx.

NI-DAQmx Help

Information about creating applications that use the NI-DAQmx driver.

LabVIEW Help

LabVIEW programming concepts and reference information about NI-DAQmx VIs and 
functions.

NI-DAQmx C Reference Help

Reference information for NI-DAQmx C functions and NI-DAQmx C properties.

NI-DAQmx .NET Help Support for Visual Studio

Reference information for NI-DAQmx .NET methods and NI-DAQmx .NET properties, key 
concepts, and a C enum to .NET enum mapping table.

Summary of Contents for PXIe-4310

Page 1: ...ce Temperature Changes 6 Accuracy Verification 6 Adjustment 8 Reverification 9 Specifications 9 Worldwide Support and Services 9 Software Calibrating the PXIe 4310 requires the installation of NI DAQm...

Page 2: ...nd reference information PXIe 4310 Specifications PXIe 4310 specifications and calibration interval NI DAQmx Readme Operating system and application software support in NI DAQmx NI DAQmx Help Informat...

Page 3: ...tute using the requirements listed in Table 1 Table 1 Recommended Equipment for PXIe 4310 Verification Equipment Recommended Model Requirements Calibrator Fluke 5700A If this instrument is unavailable...

Page 4: ...B 4310 10V Analog Signal Names Signal Names Reference Direction Description COM 0 7 Analog Input Isolated Ground These terminals are the reference point for differential analog input measurements AI 0...

Page 5: ...tures and temperature drift Allow a warm up time of at least 15 minutes to ensure that the PXIe 4310 measurement circuitry is at a stable operating temperature Ensure that the PXI PXI Express chassis...

Page 6: ...ettings tab Temperature information is displayed in the Self Calibration and External Calibration sections of the Settings tab Accuracy Verification This section provides instructions for verifying th...

Page 7: ...acquisition task 9 For each value in Table 4 repeat steps 3 through 8 10 Disconnect the calibrator from the device You have finished verifying the analog input specification on your device Note If ver...

Page 8: ...le referenceVoltage 7 5 Table 4 PXIe 4310 Analog Input Values Range V Test Point 24 Hour Limits 2 Year Limits Min Max Location Value V Lower Limit V Upper Limit V Lower Limit V Upper Limit V 10 10 Pos...

Page 9: ...tment verify that you have met the Test Conditions before returning your device to NI Refer to Worldwide Support and Services for assistance in returning the device to NI Specifications Refer to the P...

Page 10: ...at ni com patents You can find information about end user license agreements EULAs and third party legal notices in the readme file for your NI product Refer to the Export Compliance Information at n...

Reviews: