National Instruments PXIe-4302 Calibration Procedure Download Page 11

PXIe-4302/4303 and TB-4302C Calibration Procedure

|

© National Instruments Corporation

|

11

Accuracy Verification

The TB-4302C has a total of 32, 5 

 shunt resistors, one for each channel. The reference 

designators of the shunt resistors range from R10 to R41 as shown in Figure 3.

Figure 3.  

TB-4302C Circuit Board Shunt Resistor Locator Diagram

1

 R10, R11, R12, R13, R14, R15, R16, R17
(Bottom to Top)

2

 R21, R20, R19, R18, R25, R24, R23, R22
(Bottom to Top)

3

 R26, R27, R28, R29, R30, R31, R32, R33
(Bottom to Top)

4

 R37, R36, R35, R34, R41, R40, R39, R38
(Bottom to Top)

FOR PATENTS: NI.COM/PATENTS

AI3

VSUP

AI2

VSUP

AI4

AI5

AI7

VSUP

AI6

VSUP

VSUP

VSUP

AI0

AI1

VSUP

VSUP

AI2

VSUP

AI1

VSUP

AI3

AI4

AI15

VSUP

AI8

VSUP

VSUP

VSUP

A9

AI0

VSUP

VSUP

AI19

VSUP

AI18

VSUP

AI20

AI21

AI23

VSUP

AI22

VSUP

VSUP

VSUP

AI16

AI17

VSUP

VSUP

AI28

VSUP

AI27

VSUP

AI29

AI30

AI31

VSUP

AI24

VSUP

VSUP

VSUP

A25

AI26

VSUP

VSUP

S/N

158141A-01L

NI TB-4302C

GND

PFI0

AISENSE

AISENSE

AISENSE

AISENSE

AISENSE

AISENSE

AISENSE

AISENSE

AISENSE

AISENSE

AISENSE

AISENSE

AISENSE

AISENSE

AISENSE

AISENSE

AISENSE

AIGND

AISENSE

VSUP

J2

J4

J3

J7

J5

J9

J21

J1

9

J1

0

J6

J1

5

J1

1

J1

7

U9

U1

1

U1

0

J1

2

J8

J1

3

J1

8

J1

4

CR3

R37

RT28

R36

RT27

R35

RT26

R34

RT25

R41

RT32

R40

RT31

R39

RT30

R38

RT29

RT

17

R26

RT

18

R27

RT

19

R28

RT20

R29

RT21

R30

RT24

R33

RT23

R32

RT22

R31

RT

1

R1

0

RT

2

R1

1

RT

3

R1

2

RT

4

R1

3

RT

5

R1

4

RT

6

R1

5

RT

7

R1

6

RT

8

R1

7

R24

RT

15

R23

RT

14

R22

RT

13

R1

9

R

18

R25

RT

10

RT

9

R21

R

20

RT

12

RT

11

RT

16

C6

C5

C3

C1

C2

U3

U2

R3

R1

R6

R5

R4

U4

R7

R8

CR1

U1

R2

C4

CR2

8

1

8

1

8

1

8

1

8

1

8

1

8

111

8

J1

C1

B1

A1

COPYRIGHT 2014

1

2

3

4

Summary of Contents for PXIe-4302

Page 1: ...date 9 Reverification 10 TB 4302C Verification 10 Test Equipment 10 Accuracy Verification 11 Specifications 14 Revision History 14 NI Services 14 Software Calibrating the PXIe 4302 4303 requires you t...

Page 2: ...ting the PXIe 4302 4303 Test Equipment Table 1 lists the equipment recommended for the performance verification and adjustment procedures of the PXIe 4302 4303 If the recommended equipment is not avai...

Page 3: ...SMU PXIe 4139 Noise 0 1 Hz to 10 Hz peak to peak is 60 V or better at 10 V Noise 0 1 Hz to 10 Hz peak to peak is 2 V or better at 100 mV Table 1 Recommended Equipment for PXIe 4302 4303 Verification a...

Page 4: ...th 1 or better tolerance to build a voltage divider to bias the PXIe 4139 output and set the common mode input of the PXIe 4302 4303 to zero volts Connect one resistor between AI and AIGND and the oth...

Page 5: ...n Guide and Terminal Block Specifications for information about how to install the software and hardware and how to configure the device in Measurement Automation Explorer MAX Note When a device is co...

Page 6: ...Voltage output Sense Remote Range 600 mV range for test points less than 100 mV 60 V range for all other test points Current Limit 20 mA Current Limit Range 200 mA Table 4 PXI 4071 Voltage Measurement...

Page 7: ...onfiguration Value Channel Name Dev1 ai0 32 Task AI voltage Sample Mode Finite samples Sample Clock Rate 5000 Samples per Channel 5000 Maximum Value Appropriate maximum range value from Table 6 Minimu...

Page 8: ...with the following parameters Device In Dev1 Password NI1 4 Call the 4302 4303 instance of the DAQmx Setup SC Express Calibration function with the following parameters calhandle in calhandle output f...

Page 9: ...eps 5 through 8 for the remaining Test Point values from Table 7 for the corresponding range that was configured in step 4 10 Repeat steps 4 through 9 for the remaining ranges from Table 7 11 Call the...

Page 10: ...returning the device to NI TB 4302C Verification This section provides information for verifying the performance of the TB 4302C Test Equipment Table 8 lists the equipment recommended for verifying t...

Page 11: ...VSUP VSUP A9 AI0 VSUP VSUP AI19 VSUP AI18 VSUP AI20 AI21 AI23 VSUP AI22 VSUP VSUP VSUP AI16 AI17 VSUP VSUP AI28 VSUP AI27 VSUP AI29 AI30 AI31 VSUP AI24 VSUP VSUP VSUP A25 AI26 VSUP VSUP S N 158141A 0...

Page 12: ...annel to Shunt Reference Designator Correlation Channel Shunt Reference Designator CH0 R10 CH1 R11 CH2 R12 CH3 R13 CH4 R14 CH5 R15 CH6 R16 CH7 R17 CH8 R21 CH9 R20 CH10 R19 CH11 R18 CH12 R25 CH13 R24 C...

Page 13: ...1 6 Compare the results to the Lower Limit and Upper Limit values in Table 11 If the results are between these values the device passes the test 7 Repeat steps 3 through 6 for all other 5 shunt resist...

Page 14: ...Customers The data contained in this manual was developed at private expense and is subject to the applicable limited rights and restricted data rights as set forth in FAR 52 227 14 DFAR 252 227 7014...