National Instruments Corporation
9
NI TB-2708 Installation Guide
Specifications
This section lists the specifications of the TB-2708. These specifications
are typical at 25 °C unless otherwise specified.
Input/Output
Refer to the documentation for your PXI-6115/6120 S Series device to
determine the input/output specifications for your application.
Caution
Do
not
apply an input voltage greater than 42 V
pk
/60 VDC to the TB-2708. Input
voltages greater than 42 V
pk
/60 VDC can damage the TB-2708 and any device connected
to it, including the host computer. Overvoltage can also cause an electric shock hazard for
the operator. National Instruments is not liable for damage or injury resulting from such
misuse.
Physical
Dimensions
Height.............................................. 10.7
×
8.6
×
2.0 cm
(4.2
×
3.4
×
0.8 in.)
Weight............................................. 144.6 g (5.1 oz)
I/O connectors
68-position SCSI-II type ................. 1, female
SMB jacks....................................... 9
14-pin auxiliary connector .............. 1, Molex 14-pin microfit
connector
Shock and Vibration
Operational shock .................................. 30 g peak, half-sine, 11 ms pulse
(Tested in accordance with
IEC-60068-2-27. Test profile
developed in accordance with
MIL-PRF-28800F.)
Random vibration
Operating ........................................ 5 to 500 Hz, 0.3 g
rms
Nonoperating .................................. 5 to 500 Hz, 2.4 g
rms
(Tested in accordance with
IEC-60068-2-64. Nonoperating
test profile exceeds
the requirements of
MIL-PRF-28800F, Class 3.)
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com