Differential Signals
To connect differential signals, determine the type of signal source you are
using: a floating signal source or a ground-referenced signal source. Refer
to the
Differential Connection Considerations
sections of Chapter 2,
, for more information.
To measure a floating signal source, move the switch to the FS position. To
measure a ground-referenced signal source, move the switch to the GS
position. Figure A-74 shows the source type switch locations on the front
panel of the BNC DAQPads.
Figure A-74.
BNC DAQPad Front Panel
Figure A-75 shows the analog input circuitry on BNC DAQPads. When the
switch is in the FS position, AI
x
– is grounded through a 0.1
µ
F capacitor
in parallel with a 5 k
Ω
resistor.
Figure A-75.
Analog Input Circuitry
FS
GS
FS