Caution

 

Apply external signals only while the NI 5661 is powered on. Applying

external signals while the device is powered off may cause damage.

1.

Disconnect any signals from the module front panels.

2.

Power off the chassis.

3.

Remove the module from the chassis and inspect it for damage. Do not reinstall a
damaged device.

4.

Reinstall the module in a different chassis slot.

5.

Power on the chassis.

6.

Verify that the device appears in MAX.

7.

Reset the device in MAX and perform a self-test.

What Should I Do if the NI 5661 Doesn't Appear in MAX?

1.

In the MAX configuration tree, click 

Devices and Interfaces

.

2.

Expand the 

Chassis

 tree to see the list of installed devices, and press <F5> to refresh the

list.

3.

If the module is still not listed, power off the system, ensure that all hardware is correctly
installed, and restart the system.

4.

Navigate to the Device Manager.

Operating System Description

 

 

Windows 8

Right-click the Start screen, and select 

All apps

»

Control Panel

»

Hardware and Sound

»

Device Manager

.

Windows 7

Select 

Start

»

Control Panel

»

Device Manager

.

Windows Vista

Select 

Start

»

Control Panel

»

System and Maintenance

»

Device

Manager

.

Windows XP

Select 

Start

»

Control Panel

»

System

»

Hardware

»

Device

Manager

.

5.

If you are using a PXI controller, verify that a 

National Instruments

 entry appears in the

system device list. Reinstall NI-RFSA and the device if error conditions appear in the list.
If you are using an MXI controller, right-click 

PCI-to-PCI Bridge

, and select 

Properties

from the shortcut menu to verify that the bridge is enabled.

What Should I Do if the Module Fails the Self-Test?

1.

Restart the system.

2.

Launch MAX, and perform the self-test again.

3.

Power off the chassis.

4.

Reinstall the failed module in a different slot.

5.

Power on the chassis.

6.

Perform the self-test again.

NI PXI-5661 Getting Started Guide

  | 

© National Instruments

  | 

19

Summary of Contents for PXI-5661

Page 1: ...142 IF digitizer module Caution The protection provided by this product may be impaired if it is used in a manner not described in this document Hot Surface If the NI 5661 has been in use it may exceed safe handling temperatures and cause burns Allow the NI 5661 to cool before removing it from the chassis Contents Electromagnetic Compatibility Guidelines 2 Verifying the System Requirements 2 Unpac...

Page 2: ... the specified EMC performance operate this product only with shielded cables and accessories Caution To ensure the specified EMC performance the length of all I O cables must be no longer than 3 m 10 ft Verifying the System Requirements To use the NI RFSA instrument driver your system must meet certain requirements For more information about minimum system requirements recommended system and supp...

Page 3: ...5B 01 4 Semi Flexible SMA to SMB Cable Part Number 192973B 01 5 Semi Flexible SMA to SMA Cable Part Number 150611A 01 6 Cable Wrench x2 7 Screwdriver Part Number 772006 01 8 Driver Software DVD 9 Read Me First Safety and Electromagnetic Compatibility 10 Maintain Forced Air Cooling Note to Users 11 NI PXI 5661 Getting Started Guide This Document NI PXI 5661 Getting Started Guide National Instrument...

Page 4: ...661 in meets the following specifications Operating ambient temperature IEC 60068 2 1 IEC 60068 2 2 0 C to 50 C Operating relative humidity IEC 60068 2 56 10 to 90 noncondensing Maximum altitude 2 000 m 800 mbar at 25 C ambient temperature Pollution Degree 2 Indoor use only Caution Clean the hardware with a soft nonmetallic brush Make sure that the hardware is completely dry and free from contamin...

Page 5: ...lot pins on the chassis backplane for any bends or damage prior to installation Do not install a module if the backplane is damaged 4 Remove the black plastic connectors from all the captive screws on the module front panel 5 Identify a supported slot in the chassis The following figure shows the symbols that indicate the slot types Figure 2 Chassis Compatibility Symbols NI PXIe 1062Q 1 2 3 4 5 1 ...

Page 6: ...l stability and also electrically connects the front panel to the chassis which can improve the signal quality and electromagnetic performance 11 Cover all empty slots using filler panels or slot blockers to maximize cooling air flow 12 Power on the chassis Related Information Installing the Software on page 4 Interconnecting the NI 5661 Modules Caution The signal pins of this product s input outp...

Page 7: ...le connect the lower of the two 10 MHz OUT connectors on the NI 5600 front panel to the CLK IN connector on the NI 5142 front panel 3 Optional Using the semi flexible SMA to SMA coaxial cable center pin connect the bidirectional PXI 10 MHz I O connector on the NI 5600 front panel to the other 10 MHz OUT connector on the NI 5600 front panel This connection drives the PXI 10 MHz backplane with the N...

Page 8: ...oting on page 18 If an issue persists after you complete a troubleshooting procedure contact NI technical support or visit ni com support Direct Connections to the NI 5600 The NI 5600 is a precision RF instrument that is sensitive to ESD and transients Ensure you are making proper direct connections to the NI 5600 to avoid damaging the device Caution Apply external signals only while the NI 5600 i...

Page 9: ... MHz 25 MHz 20 dBm MAX 30 dBm MAX WITH ATTN PXI 10 MHz I O 50 Ω 23 dBm MAX 5 V DC MAX 10 MHz OUT 1 V pp 50 Ω FREQ REF IN 50 Ω 16 dBm MAX 5 V DC MAX NI 5600 POWER STATUS RF DOWNCONVERTER Table 1 Device Front Panel Icon Definition Refer to the user documentation for required maintenance measures to ensure user safety and or preserve the specified EMC performance NI PXI 5661 Getting Started Guide Nat...

Page 10: ...he lower 10 MHz OUT connector to the REF CLK IN connector on the NI 5620 module front panel or to the CLK IN connector on the NI 5142 PXI 10 MHz I O Provides bidirectional connection to the PXI 10 MHz backplane clock You can use this connector to drive the PXI 10 MHz backplane clock only when the NI 5600 is installed in PXI Slot 2 You can use this connector to export the PXI 10 MHz backplane clock...

Page 11: ...s detected a problem with a power rail GREEN The module is functional and receiving power STATUS Indicates the status of the NI 5600 RF downconverter module PLLs OFF The module is in an uninitialized state or the module PLLs are attempting to lock GREEN The module is in a ready state applicable PLLs are locked Related Information Refer to the NI RF Vector Signal Analyzers Help for more information...

Page 12: ...f the NI 5142 IF digitizer module contains six connectors and two LEDs Figure 6 NI 5142 IF Digitizer Module Front Panel TRIG CH 0 NI PXI 5142 ACCESS ACTIVE CH 1 AUX I O 5 V MAX CLK IN CLK OUT 100 MS s OSP Digitizer 12 ni com NI PXI 5661 Getting Started Guide ...

Page 13: ...This input terminal is not used for NI 5661 operation CH 0 is the only channel with digital downconverter DDC capabilities TRIG This connector is the input terminal for a digital trigger from an external source CLK IN Connects to the lower 10 MHz OUT connector on the NI 5600 RF downconverter module front panel This connector is the input terminal for the NI 5600 internal Reference Clock signal CLK...

Page 14: ... module has detected an error The LED remains red until the error condition is removed Related Information Refer to the NI High Speed Digitizers Help for more information about triggering Configuring the NI 5661 in MAX Use Measurement Automation Explorer MAX to configure your National Instruments hardware MAX informs other programs about which devices reside in the system and how they are configur...

Page 15: ...ponent drop down listbox For the NI 5661 you must associate the NI 5142 IF digitizer module with the NI 5600 RF downconverter c Click Save in the MAX toolbar Note Module associations may be lost when you move the modules to different chassis slots 6 Self test the device modules by selecting the modules in the configuration tree and clicking Self Test in the MAX toolbar Repeat this step for all mod...

Page 16: ...ogramming the NI 5661 You can acquire data interactively using the NI RFSA Soft Front Panel SFP or use NI RFSA to program your device in the ADE of your choice 16 ni com NI PXI 5661 Getting Started Guide ...

Page 17: ...mples located in the NIDocDir NI RFSA examples directory where NIDocDir is one of the following directories Windows 8 7 Vista Users Public Public Documents National Instruments Windows XP Documents and Settings All Users Documents National Instruments You can modify an NI RFSA example to create an application with Microsoft Visual C C and all required include and library files are automatically ad...

Page 18: ...leshooting procedure contact NI technical support or visit ni com support Installation What Should I Do if the NI 5600 RF Downconverter Module Front Panel POWER LED is Off When the PXI Chassis is On If the POWER LED fails to light after you power on the PXI chassis a problem may exist with the PXI power rail a hardware module or the LED 1 Disconnect any signals from the PXI module front panel 2 Po...

Page 19: ...the system 4 Navigate to the Device Manager Operating System Description Windows 8 Right click the Start screen and select All apps Control Panel Hardware and Sound Device Manager Windows 7 Select Start Control Panel Device Manager Windows Vista Select Start Control Panel System and Maintenance Device Manager Windows XP Select Start Control Panel System Hardware Device Manager 5 If you are using a...

Page 20: ...h typically greater than 10 dBm 3 Verify the external source signal falls within the specified frequency accuracy 4 If you programmed the RF vector signal analyzer to use the NI 5600 RF downconverter module internal timebase verify the following items a Verify that one or more of the local oscillators is phase locked if the RF vector signal analyzer measures the frequency incorrectly or detects no...

Page 21: ...ront panel interconnections for your RF vector signal analyzer 2 Verify the MAX device association for your NI 5661 3 If you have verified the interconnections and associations of your device and you are still unable to launch the NI RFSA SFP try uninstalling and then reinstalling NI RFSA Related Information Interconnecting the NI 5661 Modules on page 6 Configuring the NI 5661 in MAX on page 14 Me...

Page 22: ...ver software EXPLORE LEARN CREATE DISCOVER RF Solutions ni com rf Services ni com services NI Community ni com community Located at ni com gettingstarted Located at ni com manuals Located using the NI Example Finder Tip The NI RF Vector Signal Analyzers Help is an HTML version of a traditional user manual that includes detailed information about RF fundamentals device features and programming with...

Page 23: ...If your product supports calibration you can obtain the calibration certificate for your product at ni com calibration National Instruments corporate headquarters is located at 11500 North Mopac Expressway Austin Texas 78759 3504 National Instruments also has offices located around the world For telephone support in the United States create your service request at ni com support or dial 1 866 ASK ...

Page 24: ... and third party legal notices in the readme file for your NI product Refer to the Export Compliance Information at ni com legal export compliance for the National Instruments global trade compliance policy and how to obtain relevant HTS codes ECCNs and other import export data NI MAKES NO EXPRESS OR IMPLIED WARRANTIES AS TO THE ACCURACY OF THE INFORMATION CONTAINED HEREIN AND SHALL NOT BE LIABLE ...

Reviews: