National Instruments Corporation
55
NI PXI-4070 Calibration Procedure
•
Input Resistance
=
NIDMM_VAL_RESISTANCE_NA
•
Expected Value
=
–0.02
10. Output 0 A on the calibrator with the current output set to
AUX
.
11. Call
niDMM_CalAdjustOffset
with the following parameters:
•
Mode
=
NIDMM_VAL_DC_CURRENT
•
Range
=
0.02
•
Input Resistance
=
NIDMM_VAL_RESISTANCE_NA
12. Call
niDMM_CalAdjustOffset
with the following parameters:
•
Mode
=
NIDMM_VAL_AC_CURRENT
•
Range
=
0.01
•
Input Resistance
=
NIDMM_VAL_RESISTANCE_NA
13. Output 200 mA on the calibrator with the current output set to
AUX
.
14. Call
niDMM_CalAdjustGain
with the following parameters:
•
Mode
=
NIDMM_VAL_DC_CURRENT
•
Range
=
0.2
•
Input Resistance
=
NIDMM_VAL_RESISTANCE_NA
•
Expected Value
=
0.2
15. Output –200 mA on the calibrator with the current output set to
AUX
.
16. Call
niDMM_CalAdjustGain
with the following parameters:
•
Mode
=
NIDMM_VAL_DC_CURRENT
•
Range
=
0.2
•
Input Resistance
=
NIDMM_VAL_RESISTANCE_NA
•
Expected Value
=
–0.2
17. Output 0 A on the calibrator with the current output set to
AUX
.
18. Call
niDMM_CalAdjustOffset
with the following parameters:
•
Mode
=
NIDMM_VAL_DC_CURRENT
•
Range
=
0.2
•
Input Resistance
=
NIDMM_VAL_RESISTANCE_NA
19. Call
niDMM_CalAdjustOffset
with the following parameters:
•
Mode
=
NIDMM_VAL_AC_CURRENT
•
Range
=
0.1
•
Input Resistance
=
NIDMM_VAL_RESISTANCE_NA
20. Output 1 A on the calibrator with the current output set to
AUX
.