Storage Environment
Ambient temperature range
-40 to 71 °C
(Tested in accordance with IEC 60068-2-1 and
IEC 60068-2-2. Meets MIL-PRF-28800F
Class 3 limits.)
Relative humidity range
5 to 95%, noncondensing
(Tested in accordance with IEC 60068-2-56)
Shock and Vibration
Operational shock
30 g peak, half-sine, 11 ms pulse
(Tested in accordance with IEC 60068-2-27.
Test profile developed in accordance with
MIL-PRF-28800F.)
Random Vibration
Operating
5 to 500 Hz, 0.3 g
rms
Nonoperating
5 to 500 Hz, 2.4 g
rms
(Tested in accordance with IEC 60068-2-64.
Nonoperating test profile exceeds the
requirements of MIL-PRF-28800F, Class 3.)
Safety
This product is designed to meet the requirements of the following standards of safety for
electrical equipment for measurement, control, and laboratory use:
•
IEC 61010-1, EN 61010-1
•
UL 61010-1, CSA 61010-1
Note
For UL and other safety certifications, refer to the product label or the
section.
Electromagnetic Compatibility
This product is designed to meet the requirements of the following standards of EMC for
electrical equipment for measurement, control, and laboratory use:
•
EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity
•
EN 55011 (CISPR 11): Group 1, Class A emissions
•