INSTALLATION INSTRUCTIONS

NI TB-2636

4

×

32 2-Wire Terminal Block for the NI PXI-2529

This document describes how to install and connect signals to the 
National Instruments TB-2636 terminal block. Refer to the 

NI Switches 

Getting Started Guide

 to determine when to install the NI TB-2636.

Introduction 

The NI TB-2636 terminal block installs in front of the NI PXI-2529 switch 
module and has screw terminals that provide access to the rows and 
columns of the matrix. 

Caution

This terminal block is rated for Measurement Category I and intended to carry 

signal voltages no greater than 150 V. This module can withstand up to 800 V impulse 
voltage. Do 

not

 use this module for connection to signals or for measurements within 

Categories II, III, or IV. Do 

not

 connect to MAINS supply circuits (for example, wall 

outlets) of 115 or 230 VAC. Refer to the 

Read Me First: Safety and Radio-Frequency 

Interference

 document for more information on measurement categories.

When module terminals are hazardous voltage LIVE (> 42.4 Vpk/60 VDC), you must 
ensure that devices and circuits connected to the module are properly insulated from 
human contact.

Conventions

The following conventions are used in this document:

»

The 

»

 symbol leads you through nested menu items and dialog box options 

to a final action. The sequence 

File»Page Setup»Options

 directs you to 

pull down the 

File

 menu, select the 

Page Setup

 item, and select 

Options

 

from the last dialog box.

̭͈Ρ΅νιϋΠ͉ͅȂ඾ུࢊβȜΐ͜܄̞̳ͦ̀͘͘ȃ

Summary of Contents for NI TB-2636

Page 1: ... 800 V impulse voltage Do not use this module for connection to signals or for measurements within Categories II III or IV Do not connect to MAINS supply circuits for example wall outlets of 115 or 230 VAC Refer to the Read Me First Safety and Radio Frequency Interference document for more information on measurement categories When module terminals are hazardous voltage LIVE 42 4 Vpk 60 VDC you mu...

Page 2: ...the proper names of disk drives paths directories programs subprograms subroutines device names functions operations variables filenames and extensions 1 Unpack the Terminal Block The terminal block is shipped in an antistatic package to prevent electrostatic discharge ESD that can damage several components on the terminal block To avoid such damage when you handle the terminal block take the foll...

Page 3: ...y of space for the signal cables 4 Pull the signal wires through the strain relief opening shown in Figure 1 5 Connect the safety earth ground to the safety ground lug 6 Connect the wires to the terminals by inserting the stripped end of the wire into the terminal Secure the wire by tightening the screw for each terminal No bare wire should extend past the screw terminal Exposed wire increases the...

Page 4: ...igure 1 NI TB 2636 Terminal Block 1 Strain Relief Bar 2 Strain Relief Opening 3 Signal Wires 4 Strain Relief Screw 5 Jackscrew 6 Safety Ground Lug 7 Terminal Block Top Cover 8 Top Cover Screw 9 Rear Connector 10 Screw Terminals 1 2 5 4 7 6 10 9 8 3 ...

Page 5: ...FETY EARTH GROUND TO THE SAFETY GROUND LUG NI TB 2636 Terminal Reference Guide 100 POSITION CONNECTOR C12 C12 C13 C13 C14 C14 C15 C15 C28 C28 C29 C29 C30 C30 C31 C31 C8 C8 C9 C9 C10 C10 C11 C11 C24 C24 C25 C25 C26 C26 C27 C27 C20 C20 C21 C21 C22 C22 C23 C23 C16 C16 C17 C17 C18 C18 C19 C19 C4 C4 C5 C5 C6 C6 C7 C7 R0 R0 R1 R1 R2 R2 R3 R3 C0 C0 C1 C1 C2 C2 C3 C3 ...

Page 6: ...lowing steps while referring to Figure 3 1 Connect the NI PXI 2529 front connector to its mating connector on the terminal block 2 Tighten the top and bottom jackscrews on the terminal block to hold it securely in place Do not overtighten the screws Figure 3 Installing the NI TB 2636 Terminal Block 1 Jackscrew 2 Front Connector 3 NI PXI 2529 4 NI TB 2636 3 4 2 1 ...

Page 7: ... EN 55011 Emissions Group 1 Class A CE C Tick ICES and FCC Part 15 Emissions Class A Note For EMC compliance operate this device according to product documentation CE Compliance This product meets the essential requirements of applicable European Directives as amended for CE marking as follows 73 23 EEC Low Voltage Directive safety 89 336 EEC Electromagnetic Compatibility Directive EMC Note Refer ...

Page 8: ...tional Instruments trademarks Other product and company names mentioned herein are trademarks or trade names of their respective companies For patents covering National Instruments products refer to the appropriate location Help Patents in your software the patents txt file on your CD or ni com patents 2007 National Instruments Corporation All rights reserved 374615B Nov07 ...

Page 9: ...は 150 V 以下で 最大 800 V インパルス電圧に対して耐性があります Category II III または IV の信号を このモジュールに接続したり測定しない でください 115 または 230 VAC のコンセントを使用する電源回路 例 壁コン セント に接続しないでください 測定カテゴリの詳細については はじめに お読みください 安全対策と無線周波数妨害について を参照してください モジュール端子が危険電圧 42 4 Vpk 60 VDC で活電状態の場合は デバイスとモジュールに接続されている回路が人体に触れないよう必ず適 切に絶縁してください 表記規則 このドキュメントでは 以下の表記規則を使用します 矢印 は ネスト化されたメニュー項目やダイアログボックスのオプ ションをたどっていくと目的の操作項目を選択できることを示します たとえば ファイル ページ設定 オプショ...

Page 10: ...コードの一部 プログラムサンプル 構文例を表します また ディスクドライブ パス ディレクトリ プログラム サブプログラム サブルーチンなどの名称 デバイス名 関数 操作 変数 ファイル名お よび拡張子の引用にも使用されます 1 端子台を箱から取り出す 端子台は 部品に破損をもたらす静電放電 ESD を防止するために静電 気防止用パッケージに包装されて出荷されています このような破損を回 避するには 以下の予防措置を行ってください 注意 露出しているコネクタピンには絶対に触れないでください 接地ストラップを使用したり 接地された物体に触れて 身体を接地 する 静電気防止用パッケージをシャーシの金属部分に接触させてから 端子台をパッケージから取り出す 端子台を箱から取り出し 部品がゆるんでいないかどうか また 破損箇 所がないかどうか調べます 端子台が破損している場合は ナショナルイ ンスツ...

Page 11: ...バーのネジを外しま す 2 端子台から上部カバーを注意して取り外します 3 ストレインリリーフバーの 2 つのネジを緩めて 信号ケーブルを通 せるようにスペースを空けます 4 図 1 に示されているように ワイヤをストレインリリーフバー間の 隙間に通します 5 アースを安全接地用のつまみに接続します 6 ワイヤの端の裸線を端子に差し込んで端子に接続します 各端子に付 いているネジを締めてワイヤを固定します 裸線がネジ端子の外に露 出しないように注意してください ワイヤが露出しているとショート する可能性があり 正しく作動しない場合があります 7 ストレインリリーフバー間に通されている端子台のワイヤを引っ張っ て たるみを解消します 8 ワイヤが固定されるまで ストレインリリーフバーにある 2 つのネ ジを締めます 9 端子台の上部カバーを元のように取り付けます 10 上部カバーのネジを締め...

Page 12: ...NI TB 2636 取り付け手順 4 ni com jp 図 1 NI TB 2636 端子台 1 ストレインリリーフバー 2 ストレインリリーフバー間の隙間 3 信号線 4 ストレインリリーフ用ネジ 5 ジャックネジ 6 安全接地用つまみ 7 端子台の上部カバー 8 上部カバーのネジ 9 後部コネクタ 10 ネジ留め式端子 1 2 5 4 7 6 10 9 8 3 ...

Page 13: ...C15 C15 C28 C28 C29 C29 C30 C30 C31 C31 C8 C8 C9 C9 C10 C10 C11 C11 C24 C24 C25 C25 C26 C26 C27 C27 C20 C20 C21 C21 C22 C22 C23 C23 C16 C16 C17 C17 C18 C18 C19 C19 C4 C4 C5 C5 C6 C6 C7 C7 R0 R0 R1 R1 R2 R2 R3 R3 C0 C0 C1 C1 C2 C2 C3 C3 㧗㟁ᅽ䛜Ꮡᅾ䛩䜛ሙྜ䛻䛿䚸䜰䞊䝇䜢 Ᏻ ᥋ᆅ 䛴䜎䜏䛻᥋ 䛧䛶䛟䛰䛥䛔䚹 㻝㻜㻜䝢䞁䝁䝛䜽䝍 㻺㻵㻌㼀㻮㻙㻞㻢㻟㻢 Ꮚ䛾ཧ 䜺䜲䝗 ...

Page 14: ...TB 2636 端子台を NI PXI 2529 のフロントパネルに接続するには 図 3 を参照しながら以下の手順に従ってください 1 NI PXI 2529 のフロントコネクタを端子台のメイトコネクタに接続し ます 2 端子台の上部と下部にあるジャックネジを締めて 端子台をしっかり と固定します ネジは締めすぎないようにしてください 図 3 NI TB 2636 端子台を取り付ける 1 ジャックネジ 2 正面コネクタ 3 NI PXI 2529 4 NI TB 2636 3 4 2 1 ...

Page 15: ...N 61326 EMC 必要条件 最小イミュニティ EN 55011 エミッション Group 1 Class A CE C Tick ICES FCC Part 15 エミッション Class A メモ このデバイスは EMC 要件に適合するため 製品ドキュメントに従って操作し てください CE 適合 この製品は 以下のように CE マーク改正に基づいて 該当する EC 理 事会指令による基本的要件に適合しています 73 23 EEC 低電圧指令 安全性 89 336 EEC 電磁適合性指令 EMC メモ この製品のこのほかの適合規格については この製品の適合宣言 DoC を参照してください この製品の適合宣言を入手するには ni com certification 英語 にアクセスして型番または製品ラインで検索し 保証の欄の該当するリンクをクリックしてください 廃電気電子機器指令 WE...

Page 16: ...ents の商標の詳細については ni com legal の Terms of Use セクションを参照してください 本文書中に記載されたその他の 製品名および企業名は それぞれの企業の商標または商号です National Instruments の製品を保護 する特許については ソフトウェアに含まれている特許情報 ヘルプ 特許情報 CD に含まれている patents txt ファイル または ni com patents のうち 該当するリソースから参照してください 2007 National Instruments Corporation All rights reserved 374615B 01 2007 年 11 月 ...

Reviews: