A-2
|
ni.com
Appendix A
Specifications
Electrical
Physical
Board dimensions .............................................Four-wide 3U PXI Express module
Slot requirements ..............................................One system slot plus three controller
expansion slots
Compatibility ....................................................Fully compatible with
PXI Express
Specification
1.0
Weight ...............................................................1.32 kg (2.91 lb) typical
Environment
Maximum altitude.............................................2,000 m (800 mbar)
(at 25 °C ambient temperature)
Pollution Degree ...............................................2
Indoor use only.
Caution
Clean the PXIe-8840 with a soft nonmetallic brush. Make sure that the
device is completely dry and free from contaminants before returning it to service.
Operating Environment
Ambient temperature range ..............................5 to 50 °C
(Tested in accordance with IEC-60068-2-1 and
IEC-60068-2-2. Meets MIL-PRF-28800F
Class 3 high temperature limit.)
Extended temperature option...............................0 to 55 °C
(Tested in accordance with IEC-60068-2-1 and
IEC-60068-2-2. Meets MIL-PRF-28800F Class 3
low temperature limit and MIL-PRF-28800F
Class 2 high temperature limit.)
Voltage (V)
Current (Amps)
Typical
Maximum
+3.3 V
1.03 A
1.65 A
+5 V
1.33 A
2.48 A
+12 V
2.87 A
5.16 A
-12 V
0 A
0 A
+5 V
Aux
0.20 A
0.41 A
Note
: Does not include any attached USB devices or ExpressCard.