Physical Characteristics

Weight (unloaded)

412 g (14.53 oz)

Dimensions

3U, one-slot, PXI Express/CompactPCI Express module; 2.0 cm × 13.0 cm ×

21.6 cm (0.8 in. × 5.1 in. × 8.5 in.), nominal

Maintenance

Clean the hardware with a soft, nonmetallic brush. Make sure that the hardware is completely dry and free from contaminants before returning it to service.

Environmental

This product was tested in accordance with:
• IEC 60068-2-1
• IEC 60068-2-2
• IEC 60068-2-56

Temperature

Operating

0 °C to 55 °C

Operating humidity

10% to 70%, noncondensing. Derate 1.3% per °C above 40 °C.

Storage humidity

5% to 95%, noncondensing.

Pollution Degree

2

Maximum altitude

2,000 m (800 mbar) (at 25 °C ambient temperature)

Indoor use only.

Safety Standards

This product is designed to meet the requirements of the following electrical equipment safety standards for measurement, control, and laboratory use:
• IEC 61010-1, EN 61010-1
• UL 61010-1, CSA C22.2 No. 61010-1

Electromagnetic Compatibility

This product meets the requirements of the following EMC standards for electrical equipment for measurement, control, and laboratory use:
• EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity
• EN 55011 (CISPR 11): Group 1, Class A emissions
• AS/NZS CISPR 11: Group 1, Class A emissions
• FCC 47 CFR Part 15B: Class A emissions
• ICES-001: Class A emissions

Note

 

In the United States (per FCC 47 CFR), Class A equipment is intended for use in commercial, light-industrial, and heavy-industrial locations. In

Europe, Canada, Australia, and New Zealand (per CISPR 11), Class A equipment is intended for use only in heavy-industrial locations.

Note

 

Group 1 equipment (per CISPR 11) is any industrial, scientific, or medical equipment that does not intentionally generate radio frequency

energy for the treatment of material or inspection/analysis purposes.

Note

 

For EMC declarations and certifications, refer to the 

Online Product Certification

 section.

CE Compliance 

This product meets the essential requirements of applicable European Directives, as follows:
• 2014/35/EU; Low-Voltage Directive (safety)
• 2014/30/EU; Electromagnetic Compatibility Directive (EMC)

2

  | 

ni.com

  | 

PXIe-4143 Safety, Environmental, and Regulatory Information