1.
Call one of the following:
•
niDCPower Initialize External Calibration VI
•
niDCPower_InitExtCal function
2.
Call one of the following, specifying Commit in
calibration close action
:
•
niDCPower Close External Calibration VI
•
niDCPower_CloseExtCal function
Reverifying
Repeat the Verification section to determine the as-left status of the device. If no as-left test
limits are specified in a procedure, reuse the as-found test limits.
Note
If any test fails reverification after performing an adjustment, verify that you
have met the Test Conditions before returning the device to NI. Refer to
for more information about support resources or service
requests.
Related Information
The performance verification procedures assume that adequate traceable uncertainties are
available for the calibration references.You must complete all verification procedures in
the specified order.
Worldwide Support and Services
The National Instruments website is your complete resource for technical support. At
, you have access to everything from troubleshooting and application development
self-help resources to email and phone assistance from NI Application Engineers.
Visit
for NI Factory Installation Services, repairs, extended warranty, and
other services.
Visit
to register your National Instruments product. Product registration
facilitates technical support and ensures that you receive important information updates from
NI.
A Declaration of Conformity (DoC) is our claim of compliance with the Council of the
European Communities using the manufacturer’s declaration of conformity. This system
affords the user protection for electromagnetic compatibility (EMC) and product safety. You
can obtain the DoC for your product by visiting
calibration, you can obtain the calibration certificate for your product at
National Instruments corporate headquarters is located at 11500 North Mopac Expressway,
Austin, Texas, 78759-3504. National Instruments also has offices located around the world.
For telephone support in the United States, create your service request at
NI PXIe-4138 Calibration Procedure
|