background image

Troubleshooting

If an issue persists after you complete a troubleshooting procedure, contact NI technical
support or visit 

ni.com/support

.

What Should I Do if the NI 4112/4113 Doesn't Appear
in MAX?

1.

In the MAX configuration tree, click 

Devices and Interfaces

.

2.

Expand the 

Chassis

 tree to see the list of installed devices, and press <F5> to refresh the

list.

3.

If the module is still not listed, power off the system, ensure that all hardware is correctly
installed, and restart the system.

4.

Navigate to the Device Manager.

Operating System Description

 

 

Windows 8

Right-click the Start screen, and select 

All apps

»

Control Panel

»

Hardware and Sound

»

Device Manager

.

Windows 7

Select 

Start

»

Control Panel

»

Device Manager

.

Windows Vista

Select 

Start

»

Control Panel

»

System and Maintenance

»

Device

Manager

.

Windows XP

Select 

Start

»

Control Panel

»

System

»

Hardware

»

Device

Manager

.

5.

If you are using a PXI controller, verify that a 

National Instruments

 entry appears in the

system device list. Reinstall NI-DCPower and the device if error conditions appear in the
list. If you are using an MXI controller, right-click 

PCI-to-PCI Bridge

, and select

Properties

 from the shortcut menu to verify that the bridge is enabled.

10

  | 

ni.com

  | 

NI PXIe-4112/4113 Getting Started Guide

Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Summary of Contents for NI PXIe-4112

Page 1: ...service in house repair center WE BUY USED EQUIPMENT Sell your excess underutilized and idle used equipment We also offer credit for buy backs and trade ins www artisantg com WeBuyEquipment REMOTE IN...

Page 2: ...gramming the NI 4112 4113 9 Troubleshooting 10 What Should I Do if the NI 4112 4113 Doesn t Appear in MAX 10 Where To Go Next 11 Worldwide Support and Services 11 Electromagnetic Compatibility Guideli...

Page 3: ...ust meet certain requirements For more information about minimum system requirements recommended system and supported application development environments ADEs refer to the product readme which is ava...

Page 4: ...ense Jumpers 6 Extra Screws for the Output Connector Assembly 7 Additional Auxiliary Power Connector 8 Driver Software DVD 9 NI PXIe 4112 4113 Getting Started Guide this document 10 Maintain Forced Ai...

Page 5: ...nt you are using the NI 4112 4113 in meets the following specifications Operating Environment Ambient temperature range 0 C to 55 C Tested in accordance with IEC 60068 2 1 and IEC 60068 2 2 Meets MIL...

Page 6: ...may see access and security messages during installation Accept the prompts to complete the installation 4 When the installer completes select Restart in the dialog box that prompts you to restart sh...

Page 7: ...lation 2 3 NI PXIe 1075 1 1 Chassis 2 Hardware Module 3 Ejector Handle in Down Unlatched Position 9 Latch the module in place by pulling up on the ejector handle 10 Secure the device front panel to th...

Page 8: ...the auxiliary power input connector 13 Connect the output connector assembly to the device Tighten the thumbscrews on the output connector assembly to hold it in place Note The NI 4112 4113 ships with...

Page 9: ...to configure your National Instruments hardware MAX informs other programs about which devices reside in the system and how they are configured MAX is automatically installed with NI DCPower 1 Launch...

Page 10: ...amming Options Application Programming Interface API Location Description NI DCPower SFP Available from the start menu at Start All Programs National Instruments NI DCPower NI DCPower Soft Front Panel...

Page 11: ...ck the Start screen and select All apps Control Panel Hardware and Sound Device Manager Windows 7 Select Start Control Panel Device Manager Windows Vista Select Start Control Panel System and Maintena...

Page 12: ...y NI PXIe 4112 OR Worldwide Support and Services The National Instruments website is your complete resource for technical support At ni com support you have access to everything from troubleshooting a...

Page 13: ...of their respective companies For patents covering National Instruments products technology refer to the appropriate location Help Patents in your software the patents txt file on your media or the N...

Page 14: ...service in house repair center WE BUY USED EQUIPMENT Sell your excess underutilized and idle used equipment We also offer credit for buy backs and trade ins www artisantg com WeBuyEquipment REMOTE IN...

Reviews: