21
|
ni.
c
o
m
|
N
I 5105
Cali
br
at
ion
P
rocedu
re
Table 6.
NI 5105 Vertical Gain Accuracy Stimuli, Calibration Test Limits, and Published Specifications
Iteration
Max Input
Frequency (MHz)
Range (V)
Input Impedance
(Ohms)
Positive Input
(V)
Negative
Input (V)
Calibration
Test Limits (V)
Published
Specification (V)
1
60
0.05
1,000,000
0.0225
-
0.0225
±0.60%
±1.00%
2
60
0.20
1,000,000
0.09
-
0.09
±0.40%
±0.65%
3
60
1.00
1,000,000
0.45
-
0.45
±0.40%
±0.65%
4
60
6.00
1,000,000
2.7
-
2.7
±0.40%
±0.65%
5
60
30.00
1,000,000
13.5
-
13.5
±0.50%
±0.75%
6
60
1.00
50
0.45
-
0.45
±0.75%
±1.00%
7
24
0.05
1,000,000
0.0225
-
0.0225
±0.60%
±1.00%
8
24
0.20
1,000,000
0.09
-
0.09
±0.40%
±0.65%
9
24
1.00
1,000,000
0.45
-
0.45
±0.40%
±0.65%
10
24
6.00
1,000,000
2.7
-
2.7
±0.40%
±0.65%
11
24
30.00
1,000,000
13.5
-
13.5
±0.50%
±0.75%
12
24
1.00
50
0.45
-
0.45
±0.75%
±1.00%