3.
Insert a screwdriver into the slot on the fuse holder.
4.
Turn counterclockwise.
5.
Pull out the holder, and remove the 5
×
20 mm glass fuse.
6.
Verify the fuse is blown by measuring discontinuity across the fuse using another meter.
7.
Insert a new fuse into the holder and slide the holder back into place.
8.
Turn the fuse holder clockwise until it snaps shut.
NI 407
x
To replace the fuse, complete the following steps.
(NI PXI-4070/4071/4072)
1.
Remove all front panel connections from the NI PXI-407
x
.
2.
Power off the chassis, and remove the NI PXI-407
x
.
3.
Locate the fuse hole shown in Figure 28.
Figure 28.
Removing the Fuse from the NI PXI-407
x
4.
Insert a screwdriver into the hole.
Caution
Do
not
cover the fuse slot on the opposite side of the NI PXI-407
x
with your hand, as doing
so may cause injury.
5.
Gently press the fuse with the screwdriver until one of the fuse clamps releases.
1
Screwdriver
2
Fuse Hole
2
1