NI 5922 Calibration Procedure
|
© National Instruments
|
5
MAX
To initiate self-calibration from MAX, complete the following steps:
1.
Disconnect or disable any AC inputs to the oscilloscope.
2.
Launch MAX.
3.
Select
My System»Devices and Interfaces
.
4.
Select the device that you want to calibrate.
5.
Initiate self-calibration using one of the following methods:
•
Click
Self-Calibrate
in the upper right corner of MAX.
•
Right-click the name of the device in the MAX configuration tree and select
Self-Calibrate
from the drop-down menu.
Scope SFP
To initiate self-calibration from the Scope SFP, complete the following steps:
1.
Disconnect or disable any AC inputs to the oscilloscope.
2.
Launch the Scope SFP by selecting
Start»Programs»National Instruments»
NI-SCOPE»Scope Soft Front Panel
.
3.
Select the device you want to calibrate using the Select Device dialog box or by selecting
Edit»Change Device
.
4.
Launch the Calibration dialog box by selecting
Utility»Self Calibration
.
5.
Click
OK
to begin self-calibration.
NI-SCOPE
To self-calibrate the NI 5922 programmatically using NI-SCOPE, complete the following steps:
1.
Disconnect or disable any AC inputs to the oscilloscope.
2.
Call
niScope_init
(niScope Initialize VI) to obtain an instrument session handle. Set the
following parameters:
•
resourceName
: The device name assigned by MAX
•
IDQuery
:
NISCOPE_VAL_VI_FALSE
•
resetDevice
:
NISCOPE_VAL_VI_FALSE
•
vi
: The returned session handle that you use to identify the instrument in all
subsequent instrument driver function calls