National Instruments NI 9860 Getting Started Manual Download Page 3

built into the product if the product is damaged in any
way. If the product is damaged, return it to NI for
repair.

Safety Guidelines for Hazardous Locations

The NI 9860 is suitable for use in Class I, Division 2, Groups A,
B, C, D, T4 hazardous locations; Class I, Zone 2, AEx nA IIC T4
and Ex nA IIC T4 hazardous locations; and nonhazardous
locations only. Follow these guidelines if you are installing the
NI 9860 in a potentially explosive environment. Not following
these guidelines may result in serious injury or death.

Caution

 

Do not disconnect I/O-side wires or

connectors unless power has been switched off or the
area is known to be nonhazardous.

Caution

 

Do not remove modules unless power has

been switched off or the area is known to be
nonhazardous.

Caution

 

Substitution of components may impair

suitability for Class I, Division 2.

NI 9860 Getting Started Guide

  | 

© National Instruments

  | 

3

Summary of Contents for NI 9860

Page 1: ...GETTING STARTED GUIDE NI 9860 NI XNET Hardware Selectable Interface Fran ais Deutsch ni com manuals...

Page 2: ...Refer to the documentation for each component in the system to determine the safety and EMC ratings for the entire system Safety Guidelines Operate the NI 9860 only as described in this document Hot S...

Page 3: ...only Follow these guidelines if you are installing the NI 9860 in a potentially explosive environment Not following these guidelines may result in serious injury or death Caution Do not disconnect I O...

Page 4: ...e in Zone 2 hazardous locations in ambient temperatures of 40 C Ta 70 C If you are using the NI 9860 in Gas Group IIC hazardous locations you must use the device in an NI chassis that has been evaluat...

Page 5: ...ion about transceiver cables refer to the NI XNET CAN LIN Transceiver Cables page on ni com The NI 9860 requires an external power supply of 9 to 30 V to power the transceiver cables Supply power to t...

Page 6: ...ransceiver Cables page on ni com for mounting guidelines to limit the thermal impact to your CompactRIO system The following figure shows the NI 9860 with Port 1 connected to an NI XNET CAN HS FD Tran...

Page 7: ...NI 9860 Hardware Overview The NI 9860 is a two port C Series module based on the XNET architecture Users can choose the physical bus protocol by plugging in the corresponding transceiver cable The NI...

Page 8: ...or more information about transceiver cables refer to the NI XNET CAN LIN Transceiver Cables page on ni com Figure 2 NI 9860 Hardware Overview Ext Pwr Supply Required Controller COM Vsup _ Pwr Sup Con...

Page 9: ...host device push the connector assembly into the host receptacle until the internal latch snaps into position The latch emits an audible click when engaged To remove the NI XNET Transceiver Cable push...

Page 10: ...moving the Power Cable The NI 9860 connects to the provided power cable with an active latching connector To connect the power cable push the end of the power cable into the two position connector unt...

Page 11: ...to disengage the latch and simultaneously pull the power cable away from the module as shown in the following figure Figure 4 Removing the NI 9860 Power Cable Pull Power Cable Push Lock Ejector Down N...

Page 12: ...XNET hardware selectable interface port Port supported transceiver cable types NI XNET transceiver cables CAN LIN Note For more information about transceiver cables refer to the transceiver cable ope...

Page 13: ...quiring a connecting cable longer than 3 m 10 ft A DC mains supply is a local DC electricity supply network in the infrastructure of a site or building Total thermal dissipation on module 1 2 W max ac...

Page 14: ...erred to as MAINS voltage MAINS is a hazardous live electrical supply system that powers equipment This category is for measurements of voltages from specially protected secondary circuits Such voltag...

Page 15: ...signed to meet the requirements of the following electrical equipment safety standards for measurement control and laboratory use IEC 61010 1 EN 61010 1 UL 61010 1 CSA 61010 1 EN 60079 0 2012 EN 60079...

Page 16: ...one 2 Ex nA IIC T4 Europe DEMKO Ex nA IIC T4 Gc Environmental Refer to the installation instructions for the chassis you are using for more information about meeting these specifications Operating tem...

Page 17: ...cables attached IP 40 Operating humidity IEC 60068 2 56 10 RH to 90 RH noncondensing Storage humidity IEC 60068 2 56 5 RH to 95 RH noncondensing Pollution degree IEC 60664 2 Maximum altitude 5 000 m I...

Page 18: ...2 27 30 g 11 ms half sine 50 g 3 ms half sine 18 shocks at 6 orientations Electromagnetic Compatibility This product meets the requirements of the following EMC standards for electrical equipment for...

Page 19: ...ct meets the essential requirements of applicable European Directives as follows 2014 35 EU Low Voltage Directive safety 2014 30 EU Electromagnetic Compatibility Directive EMC 94 9 EC Potentially Expl...

Page 20: ...l environmental information refer to the Minimize Our Environmental Impact web page at ni com environment This page contains the environmental regulations and directives with which NI complies as well...

Page 21: ...u have access to everything from troubleshooting and application development self help resources to email and phone assistance from NI Application Engineers Visit ni com services for NI Factory Instal...

Page 22: ...our product at ni com calibration National Instruments corporate headquarters is located at 11500 North Mopac Expressway Austin Texas 78759 3504 National Instruments also has offices located around th...

Page 23: ...and third party legal notices in the readme file for your NI product Refer to the Export Compliance Information at ni com legal export compliance for the National Instruments global trade compliance p...

Reviews: