NI 9263 Calibration Procedure

|

© National Instruments

|

3

Test Equipment

National Instruments recommends that you use equipment in Table 1 for calibrating the 
NI 9263. If the recommended equipment is not available, select a substitute using the 
requirements listed in Table 1.

Test Conditions

The following setup and environmental conditions are required to ensure the NI 9263

 

meets 

published specifications.

Keep connections to the NI 9263 as short as possible. Long cables and wires act as 
antennas, picking up extra noise that can affect measurements.

Verify that all connections to the NI 9263 are secure.

Use shielded copper wire for all cable connections to the NI 9263. Use twisted-pair wire to 
eliminate noise and thermal offsets.

Maintain an ambient temperature of 23 ± 5 °C. The NI 9263 board temperature will be 
greater than the ambient temperature.

Keep relative humidity below 80%.

Allow a warm-up time of at least 10 minutes to ensure that the NI 9263 measurement 
circuitry is at a stable operating temperature.

Initial Setup

Complete the following steps to set up the NI 9263.
1.

Install NI-DAQmx.

2.

Make sure that the NI cDAQ-9178 power source is not connected.

3.

Insert the module into slot 8 of the NI cDAQ-9178 chassis. Leave slots 1 through 7 of the 
NI cDAQ-9178 chassis empty.

4.

Connect the NI cDAQ-9178 chassis to your host computer.

5.

Connect the power source to the NI cDAQ-9178 chassis.

6.

Launch Measurement & Automation Explorer (MAX).

7.

Right-click the device name and select 

Self-Test 

to ensure that the device is working properly.

Table 1.  

Recommended Equipment

Equipment

Recommended Model

Requirements

DMM

NI 4070 DMM 

Use a multiranging 6 1/2 digit 
DMM with voltage measurement 
accuracy of ±40 ppm.

Chassis

NI cDAQ-9178

Connection Accessory

NI 9932

Summary of Contents for NI 9263

Page 1: ...Initial Setup 3 Verification 4 Accuracy Verification 4 Adjustment 6 Accuracy Adjustment 6 EEPROM Update 6 Reverification 7 Accuracy Under Calibration Conditions 7 Worldwide Support and Services 8 Sof...

Page 2: ...ctions and Specifications NI 9263 specific information specifications and calibration interval NI DAQmx Readme Operating system and application software support in NI DAQmx LabVIEW Help LabVIEW progra...

Page 3: ...NI 9263 board temperature will be greater than the ambient temperature Keep relative humidity below 80 Allow a warm up time of at least 10 minutes to ensure that the NI 9263 measurement circuitry is...

Page 4: ...Accuracy Verification Connections 3 Set the DMM to read DC voltage in the minimum range required to cover the NI 9263 output range of 10 V 4 Acquire a sample a Create and configure an AO voltage chann...

Page 5: ...le was purchased before 2007 then use the Revision D limits Note Test limits in Table 4 are derived using the values in Table 7 Table 4 NI 9263 Revision D Test Limits Range V Test Point 1 Year Limits...

Page 6: ...C value obtained from the array of recommended calibration voltages f Wait the appropriate amount of time for the DMM measurement to settle g Read the NI 9263 output voltage measurement from the DMM h...

Page 7: ...e device to NI Accuracy Under Calibration Conditions The values in the following table are based on calibrated scaling coefficients which are stored in the onboard EEPROM The following accuracy table...

Page 8: ...ur complete resource for technical support At ni com support you have access to everything from troubleshooting and application development self help resources to email and phone assistance from NI Ap...

Reviews: