screw terminal to signals or use for measurements
within Measurement Categories II, III, or IV.
Measurement Category II is for measurements performed on
circuits directly connected to the electrical distribution system.
This category refers to local-level electrical distribution, such as
that provided by a standard wall outlet, for example, 115 V for
U.S. or 230 V for Europe.
Caution
Do not connect the NI 9239 with screw
terminal to signals or use for measurements within
Measurement Categories III or IV.
NI 9239 with BNC Safety Voltages
Connect only voltages that are within the following limits:
Isolation
Channel-to-channel
Continuous
60 VDC, Measurement
Category I
Withstand
1,000 V, verified by a 5 s
dielectric withstand test