Understanding NI 9234 Data Rates
The freq
u
ency of a master timebase (
f
M
) controls the data rate (
f
s
)
of the NI 9234. The NI 9234 incl
u
des an internal master timebase
with a freq
u
ency of 13.1072 MHz, b
u
t the mod
u
le also can accept
an external master timebase or export its own master timebase. To
synchronize the data rate of an NI 9234 with other mod
u
les that
u
se
master timebases to control sampling, all of the mod
u
les m
u
st
share a single master timebase so
u
rce. Refer to the software help
for information abo
u
t config
u
ring the master timebase so
u
rce for
the NI 9234. Visit
ni.com/info
and enter
c
s
erie
s
doc
for
information abo
u
t C Series doc
u
mentation.
The following eq
u
ation provides the available data rates of the
NI 9234:
f
s
=
where
n
is any integer from 1 to 31.
However, the data rate m
u
st remain within the appropriate data rate
range. Refer to the
section for more information
abo
u
t the data rate range. When
u
sing the internal master timebase
f
M
256
÷