NI 9225 Calibration Procedure

|

© National Instruments

|

9

6.

Calculate the phase for each channel on the NI 9225. NI recommends using the Extract 
Single Tone Information VI to calculate phase.

7.

Compare the channel-to-channel phase difference to the limits in Table 11.

8.

Repeat steps through 7 for each test point in Table 11.

9.

Disable the function generator.

10. Disconnect the function generator from the device.

If the Phase Matching verification procedure determines that the NI 9225 is outside of the limits, 
refer to 

Where to Go for Support

 for assistance in returning the device to NI.

Adjustment

The following performance adjustment procedure describes the sequence of operation required 
to adjust the NI 9225.

Accuracy Adjustment

Complete the following procedure to adjust the accuracy performance of the NI 9225.

1.

Set the calibrator to Standby mode (STBY).

2.

Connect the NI 9225 to the calibrator as shown in Figure 1.

3.

Adjust the NI 9225.

a.

Initialize a calibration session on the NI 9225. The default password is 

NI

.

b.

Input the external temperature in degrees Celsius.

c.

Call the NI 9225 get C Series adjustment points function to obtain an array of 
recommended calibration voltages for the NI 9225.

d.

Set the calibrator to a reference value determined by the array of recommended 
calibration voltages.

e.

Set the calibrator to Operate mode (OPR).

f.

Call and configure the NI 9225 adjustment function according to Table 12.

g.

Set the calibrator to Standby mode (STBY).

Table 11.  

NI 9225 Verification Test Limits for Phase Matching

Generator Frequency (kHz)

Phase Matching Channel-to-Channel Limit

1

±0.035°

20

±0.700°

Table 12.  

Adjustment Configuration

Physical Channel

Reference Value

cDAQMod8/ai

x

The actual value of the voltage output of the calibrator

Summary of Contents for NI 9225

Page 1: ...NI 9225...

Page 2: ...ation 4 Accuracy Verification 5 Gain Matching Verification 6 Phase Matching Verification 8 Adjustment 9 Accuracy Adjustment 9 EEPROM Update 10 Reverification 10 Accuracy Under Calibration Conditions 1...

Page 3: ...ons and Specifications NI 9225 specific information specifications and calibration interval NI DAQmx Readme Operating system and application software support in NI DAQmx LabVIEW Help LabVIEW programmi...

Page 4: ...for all cable connections to the NI 9225 Use twisted pairs wire to eliminate noise and thermal offsets Maintain an ambient temperature of 23 C 5 C The NI 9225 temperature will be greater than the ambi...

Page 5: ...em safety ground 4 Install the module in slot 8 of the NI cDAQ 9178 chassis Leave slots 1 through 7 of the NI cDAQ 9178 chassis empty 5 Connect the NI cDAQ 9178 chassis to your host computer 6 Connect...

Page 6: ...to Operate mode OPR 5 Acquire and average samples a Create and configure an AI voltage task according to Table 2 b Configure the task timing according to Table 3 c Start the task d Read samples from t...

Page 7: ...us of the NI 9225 1 Connect the NI 9225 to the function generator as shown in Figure 2 Figure 2 Gain Matching Verification Connections 2 Configure function generator according to Table 5 Table 4 NI 92...

Page 8: ...nel using the AmplitudeAcq value 6 Compare the channel to channel gain difference to the maximum limit of 0 25 dB 7 Disable the function generator 8 Disconnect the function generator from the device I...

Page 9: ...re samples a Create and configure an AI voltage task according to Table 9 b Configure the task timing according to Table 10 c Start the task d Read samples from the NI 9225 e Clear the task Table 8 Fu...

Page 10: ...ust the accuracy performance of the NI 9225 1 Set the calibrator to Standby mode STBY 2 Connect the NI 9225 to the calibrator as shown in Figure 1 3 Adjust the NI 9225 a Initialize a calibration sessi...

Page 11: ...ls Reverification after performing an adjustment verify that you have met the Test Conditions before returning your device to NI Refer to Where to Go for Support for assistance in returning the device...

Page 12: ...bsite is your complete resource for technical support At ni com support you have access to everything from troubleshooting and application development self help resources to email and phone assistance...

Reviews: