![background image](http://html2.mh-extra.com/html/national-instruments/ni-9218/ni-9218_getting-started_3531113015.webp)
Related reference
■
Quarter-Bridge Connections
NI 9218
NI 9984/9985
DSUB/LEMO
RC
EXC
AI
■
Connecting 120 Ω quarter bridges requires the NI 9984.
■
Connecting 350 Ω quarter bridges requires the NI 9985.
Tip
NI recommends 2 V excitation when using a NI 9984 with 120 Ω quarter
bridges and 3.3 V excitation when using the NI 9985 with 350 Ω quarter
bridges.
Related reference
■
TEDS Connections
TEDS
Enabled Sensor
NI 9218
T+
T-
Data
COM
For more information about TEDS, visit
and enter the Info Code rdteds.