
NI 9209 Calibration Procedure
|
© National Instruments
|
3
Test Equipment
Table 1 lists the equipment recommended for the performance verification and adjustment
procedures. If the recommended equipment is not available, select a substitute using the
requirements listed in Table 1.
Test Conditions
The following setup and environmental conditions are required to ensure the NI 9209
meets
published specifications.
•
Keep connections to the device as short as possible. Long cables and wires act as antennas,
picking up extra noise that can affect measurements.
•
Verify that all connections to the device are secure.
•
Use shielded copper wire for all cable connections to the device. Use twisted-pairs wire to
eliminate noise and thermal offsets.
•
Maintain an ambient temperature of 23 ± 5 °C. The device temperature will be greater than
the ambient temperature.
•
Keep relative humidity below 80%.
•
Allow a warm-up time of at least 10 minutes to ensure that the NI 9209 measurement
circuitry is at a stable operating temperature.
Initial Setup
Complete the following steps to set up the NI 9209.
1.
Install NI-DAQmx.
2.
Make sure the NI cDAQ-9178 power source is not connected.
3.
Install the module in slot 8 of the NI cDAQ-9178 chassis. Leave slots 1 through 7 of the
NI cDAQ-9178 chassis empty.
4.
Connect the NI cDAQ-9178 chassis to your host computer.
5.
Connect the power source to the NI cDAQ-9178 chassis.
Table 1.
Recommended Equipment
Equipment
Recommended
Model
Minimum Requirements
Source
Fluke 5500A
Use a source with gain error
≤
260 ppm and an
offset error
≤
70 uV.
cDAQ Chassis
NI cDAQ-9178
—
Connection
Accessory
NI 9923
—