NI 9207 Calibration Procedure

|

© National Instruments

|

7

Adjustments

The following performance adjustment procedures describe the sequence of operation required 
to adjust the NI 9207.

Voltage Accuracy Adjustment

Complete the following procedure to adjust the voltage accuracy performance of the NI 9207.
1.

Set the calibrator to Standby mode (STBY).

2.

Connect the NI 9207 to the calibrator as shown in Figure 1.

3.

Adjust the NI 9207 voltage accuracy.
a.

Initialize a calibration session on the NI 9207. The default password is 

NI

.

b.

Input the external temperature in degrees Celsius.

c.

Call the NI 9207 get C Series adjustment points function to obtain an array of 
recommended calibration voltages for the NI 9207.

d.

Set the calibrator to a reference value determined by the array of recommended 
calibration voltages.

e.

Set the calibrator to Operate mode (OPR).

f.

Call and Configure the NI 9207 gain instance of the adjustment function according to 
Table 8.

g.

Set the calibrator to Standby mode (STBY).

h.

Repeat steps d through g for each calibration voltage in the array.

i.

Disconnect the calibrator from the NI 9207.

j.

Short the voltage channel to the 9207 AIGND.

k.

Call the NI 9207 offset instance of the adjustment function for the physical channel.

l.

Close the calibration session.

4.

Remove the connections from the NI 9207.

5.

Repeat steps 1 through 4 for each voltage channel on the NI 9207.

Table 8.  

Gain Adjustment Configuration

Physical Channel

Reference Value

cDAQMod8/ai

x

The reference value from the array of adjustment points

Summary of Contents for NI-9207

Page 1: ...4 Current Accuracy Verification 5 Adjustments 7 Voltage Accuracy Adjustment 7 Current Accuracy Adjustment 8 EEPROM Update 8 Reverification 9 Accuracy Under Calibration Conditions 9 Worldwide Support a...

Page 2: ...rating Instructions and Specifications NI 9207 specific information specifications and calibration interval NI DAQmx Help Information about creating applications that use the NI DAQmx driver LabVIEW H...

Page 3: ...me of at least 10 minutes to ensure that the NI 9207 measurement circuitry is at a stable operating temperature Initial Setup Complete the following steps to set up the NI 9207 1 Install NI DAQmx 2 Ma...

Page 4: ...ode OPR 5 Acquire and average samples a Create and configure an AI voltage task on the NI 9207 according to Table 2 b Configure the AI voltage task according toTable 3 Set the ADC timing mode to high...

Page 5: ...2 Connect the NI 9207 to the calibrator as shown in Figure 2 Figure 2 Current Accuracy Connections Note If the calibrator outputs are truly floating connect the negative output to a quiet earth ground...

Page 6: ...point in Table 7 10 Disconnect the calibrator from the device 11 Repeat steps 1 through 10 for each current channel on the NI 9207 Table 5 NI 9207 Configuration for Current Accuracy Verification Input...

Page 7: ...mmended calibration voltages for the NI 9207 d Set the calibrator to a reference value determined by the array of recommended calibration voltages e Set the calibrator to Operate mode OPR f Call and C...

Page 8: ...g to Table 9 g Set the calibrator to Standby mode STBY h Repeat steps d through g for each calibration current in the array i Disconnect the calibrator from the NI 9207 Ensure that there are no connec...

Page 9: ...s which are stored in the onboard EEPROM The following accuracy table is valid for calibration under the following conditions Ambient temperature 23 C 5 C NI 9207 installed in slot 8 of an NI cDAQ 917...

Page 10: ...private expense and is subject to the applicable limited rights and restricted data rights as set forth in FAR 52 227 14 DFAR 252 227 7014 and DFAR 252 227 7015 Worldwide Support and Services The Nat...

Reviews: