22
|
ni.com
|
NI 6738/6739 Calibration Procedure
•
maxVal
:
10.0
•
units
:
DAQmx_Val_Volts
•
customScaleName
:
NULL
5.
Call
DAQmxStartTask
with the following parameter:
•
taskHandle
:
taskHandle
6.
Call
DAQmxWriteAnalogF64
with the following parameters, using a
Test Point
from
Table 7,
NI 6738/6739 Analog Output Values
•
taskHandle
:
taskHandle
•
numSampsPerChan
:
1
•
autoStart
:
1
•
timeout
:
10.0
•
dataLayout
:
DAQmx_Val_GroupByChannel
•
writeArray
:
&data
•
sampsPerChanWritten
:
&samplesWritten
•
reserved
:
NULL
7.
Compare the resulting value shown by the DMM to the upper and lower limits in Table 7,
NI 6738/6739 Analog Output Values
. If the value is between these limits, the device passes
the test.
8.
Repeat steps 6 and 7 until all of the Test Point values in Table 7,
, have been tested.
9.
Call
DAQmxClearTask
with the following parameter:
•
taskHandle
:
taskHandle
10. Disconnect the DMM from AO 0, and reconnect it to AO 1, making the connections shown
in Table 5,
.
11. Repeat steps 3 through 10 for all AO channels on the device.
12. Disconnect your DMM from the device.
You have finished verifying the analog output levels on your device.
Counter Verification
NI 6738/6739 devices have only one timebase to verify, so only Counter 0 needs to be checked.
It is not possible to adjust this timebase, so only verification can be performed.
Note
Table 8,
, is based upon the most recent edition
of the specifications document for your device. Refer to the most recent
specifications document online at
.
Notice
Do not perform analog output verification and counter verification
simultaneously, as the multiple signal connections will produce noise that can affect
measurements.
Summary of Contents for NI 6738
Page 1: ...PXIe 6738...