NI 6520/6521 User Guide
16
ni.com
(NI PXI-6521 Only)
The maximum power rating for the NI PXI-6521 relays
is 60 W or 60 VA, with a maximum current of 2 A for all channels from 0 to
55 °C, external PXI chassis ambient.
(NI PCI-6521 Only)
The maximum power rating for the NI PCI-6521 relays is
60 W or 60 VA. Figure 9 demonstrates the maximum current ratings for the
NI PCI-6521 relays from 0 to 55 °C.
Figure 9.
NI PCI-6521 Relay Power Ratings
Power-On and Power-Off Conditions
The relays on the NI 6520/6521 have the default statuses of deenergized
(control bit is low) in both power-on and power-off states. For more
information on default states of the NI 6520/6521 relays, refer to the
section and the
section.
Relay Signal Connection Example
To connect to the NI 6520/6521 signals for Form C relays, connect NC,
NO, and COM, as shown in Figure 10.
Figure 10.
Relay Signal Connection Example for Form C Relays