Chapter 3
Hardware Overview
3-6
ni.com
Figure 3-3.
Analog Trigger Block Diagram for the NI 6115/6120
In below-low-level analog triggering mode, the trigger is generated when
the signal value is less than
lowValue
, as shown in Figure 3-4.
HighValue
is unused.
Figure 3-4.
Below-Low-Level Analog Triggering Mode
In above-high-level analog triggering mode, the trigger is generated when
the signal value is greater than
highValue
,
as shown in Figure 3-5.
LowValue
is unused.
PGIA
Analog
Input
CH0
+
–
ADC
ADC
ADC
DAQ-STC
Analog
Trigger
Circuit
Mux
PGIA
Analog
Input
CH1
+
–
PGIA
Analog
Input
CH2
+
–
PGIA
Analog
Input
CH3
+
–
ADC
PFI0/TRIG1
Digital Data
AC Couple
10 k
Trigger
DAC
Trigger
DAC
highValue
lowValue
lowValue
Trigger