Table 1.
Required Equipment for Calibration
Required
Equipment
Recommended
Model(s)
Parameter
Measured
Minimum
Specifications
One digital
multimeter
(DMM), optional
second DMM
1
NI PXI-4071
All parameters
Voltage: <±200 ppm
accuracy and <30 µV
resolution
Programmable
electronic load
Agilent N3302A or
Sorensen
SLM-60-30-150
Load regulation
Constant current and
constant voltage
modes: sink at least
6 A at 10 V
100 mΩ precision
current shunt
Guildline 9230 A-30 or
Ohm Labs CS10
Current output
and measurement,
voltage and
current load
regulation
±100 ppm tolerance,
±50 ppm stability,
±5 ppm/C temperature
coefficient. Minimum
current 10 A.
Two 50 Ω resistors Vishay
PTF5650R000BZEK
Remote sense
output
0.1% tolerance, 1/8 W
1 kΩ resistor
Vishay
PTF651K0000BYEK
Remote sense
output
0.1% tolerance, 1/4 W
Test Conditions
The following setup and environmental conditions are required to ensure the NI 4113 meets
published specifications:
•
Keep cabling as short as possible. Long cables act as antennas, picking up extra noise that
can affect measurements.
•
Verify that all connections to the device, including front panel connections and screws,
are secure.
•
Ensure that the PXI chassis fan speed is set to HIGH, that the fan filters (if present) are
clean, and that the empty slots contain slot blockers and filler panels. For more
information about cooling, refer to the
Maintain Forced-Air Cooling Note to Users
document available at
1
If you make measurements using only one DMM, sequentially set up the DMM connections as
specified by the procedure steps.
NI PXIe-4113 Calibration Procedure
|
© National Instruments
|
3