Table 9.
Voltage Output and Measurement Adjustment
Level Range
Limit Range and Limit
Test Point
10 V
6 A
0.1 V
10 V
1.
Set the first specified level range, limit range, and limit on the NI 4113.
2.
Set the level on the NI 4113 to the first specified test point.
3.
Take a voltage measurement using the DMM, and take a voltage measurement using the
NI 4113.
4.
Store the values from the previous step as inputs for the niDCPower Cal Adjust VI or
function called in the following steps.
5.
Repeat the previous three steps for each test point specified in the level range.
6.
Update the measurement calibration constants by configuring and calling the niDCPower
Cal Adjust Voltage Measurement VI or
niDCPower_CalAdjustVoltageMeasurement
function.
a)
Input the DMM measurements as the
measured outputs
.
b)
Input the NI 4113 measurements as the
reported outputs
.
c)
Input the specified level range as the
range
.
7.
Update the output calibration constants by configuring and calling the niDCPower Cal
Adjust Voltage Level VI or
niDCPower_CalAdjustVoltageLevel
function.
a)
Input the DMM measurements as the
measured outputs
.
b)
Input the test points as the
requested outputs
.
c)
Input the specified level range as the
range
.
8.
If more than one level range is specified, repeat the previous steps using the values
specified in each level range.
Related Information
Connecting and Configuring Equipment for Current
Adjustment
1.
Make the necessary connections for this procedure, as shown in the following figure.
NI PXIe-4113 Calibration Procedure
|