background image

©

 National Instruments Corporation

35

NI Digital Multimeters Getting Started Guide

Figure 32.  

Removing the Fuse

4.

Remove the old fuse.

5.

Verify the fuse is blown by measuring discontinuity across the fuse using another meter.

6.

Install the new fuse.

7.

Push the fuse holder back into the housing and turn it clockwise until it tightens completely.

Appendix D: Where to Go for Support

The National Instruments Web site is your complete resource for technical support. At 

ni.com/support

 you have access to everything from troubleshooting and application development 

self-help resources to email and phone assistance from NI Application Engineers.

A Declaration of Conformity (DoC) is our claim of compliance with the Council of the European 
Communities using the manufacturer’s declaration of conformity. This system affords the user 
protection for electronic compatibility (EMC) and product safety. You can obtain the DoC for your 
product by visiting 

ni.com/certification

. If your product supports calibration, you can obtain the 

calibration certificate for your product at 

ni.com/calibration

.

National Instruments corporate headquarters is located at 11500 North Mopac Expressway, Austin, 
Texas, 78759-3504. National Instruments also has offices located around the world to help address 
your support needs. For telephone support in the United States, create your service request at 

ni.com/support

 and follow the calling instructions or dial 512 795 8248. For telephone support 

outside the United States, contact your local branch office: 

Australia 1800 300 800, Austria 43 662 457990-0, Belgium 32 (0) 2 757 0020, 
Brazil 55 11 3262 3599, Canada 800 433 3488, China 86 21 5050 9800, 
Czech Republic 420 224 235 774, Denmark 45 45 76 26 00, Finland 358 (0) 9 725 72511, 
France 01 57 66 24 24, Germany 49 89 7413130, India 91 80 41190000, Israel 972 3 6393737, 
Italy 39 02 413091, Japan 81 3 5472 2970, Korea 82 02 3451 3400, Lebanon 961 (0) 1 33 28 28, 
Malaysia 1800 887710, Mexico 01 800 010 0793, Netherlands 31 (0) 348 433 466, 
New Zealand 0800 553 322, Norway 47 (0) 66 90 76 60, Poland 48 22 3390150, 
Portugal 351 210 311 210, Russia 7 495 783 6851, Singapore 1800 226 5886, 
Slovenia 386 3 425 42 00, South Africa 27 0 11 805 8197, Spain 34 91 640 0085, 
Sweden 46 (0) 8 587 895 00, Switzerland 41 56 2005151, Taiwan 886 02 2377 2222, 
Thailand 662 278 6777, Turkey 90 212 279 3031, United Kingdom 44 (0) 1635 523545

1

NI CSM-10A/200mA

2

Fuse

3

Fuse Holder

FUSE

250V/500mA Type  F

CSM-200mA

200mA SHUNT 1

Ω

MAX

250V/200mA

CAT II FUSED

HI

LO

184200B-01

2

3

1

Summary of Contents for NI 4070

Page 1: ...manuals For the latest version of NI DMM driver software visit ni com idnet Contents Conventions 2 1 Verifying the System Requirements 3 2 Unpacking 4 3 Verifying the Kit Contents 4 Other Required Ite...

Page 2: ...cautions to take to avoid injury data loss or a system crash When this symbol is marked on a product refer to the Read Me First Safety and Radio Frequency Interference document included with the devic...

Page 3: ...isting PXI slots and PXI hybrid slots PXI Express compatible compatible modules maintain hardware and software compatibility with the exception of module local bus 1 Verifying the System Requirements...

Page 4: ...Store the DMM in the antistatic envelope when not in use USB devices Remove the USB DMM from the package and inspect it for loose components or any sign of damage Notify NI if the DMM appears damaged...

Page 5: ...e Software Before installing the DMM complete the following steps to install the software 1 Optional Install an application development environment ADE such as LabVIEW or LabWindows CVI if you are dev...

Page 6: ...efore installing the hardware Refer to the Appendix B Chassis Recommendations and System Considerations section for more information about installing the DMM in a PC or PXI chassis Installing a PXI DM...

Page 7: ...Uninstalling a PXI DMM Hot Surface During operation the metal surfaces of PXI DMMs may exceed safe handling temperatures and cause burns Allow the DMM to cool before removing it from the chassis or wh...

Page 8: ...rt the PCI DMM into an open PCI slot 4 Secure the PCI DMM with the screw provided Caution It is important to completely screw the DMM into the PCI slot for both mechanical stability and to create a so...

Page 9: ...ss slot Figure 4 PCI Express Slots Installing a USB DMM To install a USB DMM refer to Figure 5 and connect the USB cable to the PC and DMM Note The LED next to the USB connector of the device will be...

Page 10: ...e zip tie bar on the underside of the USB device and tighten around the USB cable as shown in Figure 6 Figure 6 USB Cable Strain Relief Options USB Ground Lug Optional Connect the chassis ground to th...

Page 11: ...National Instruments USB devices such as the NI USB 5133 high speed digitizer Do not apply the rubber feet if you are stacking the USB DMM on top of another National Instruments USB device For secure...

Page 12: ...print the panel mounting template PDF attached in the KnowledgeBase document Go to ni com info and enter the info code rd3233 to locate the KnowledgeBase 2 Using the template mark the bottom point and...

Page 13: ...ach device Note USB devices When you first install your USB DMM Windows will recognize a new device Click Next on any dialog boxes that appear to complete the installation After Windows recognizes new...

Page 14: ...ch operating system is compatible with your DMM 3 Verify that the DMM appears under Devices and Interfaces The DMMs appear in the NI DAQmx and Traditional NI DAQ Legacy folders in Devices and Interfac...

Page 15: ...Multimeters Help located at Start All Programs National Instruments NI DMM Documentation NI Digital Multimeters Help 5 Perform a self test on the DMM to verify installation NI DAQmx devices Right clic...

Page 16: ...ys refer to the specifications document included with the DMM before connecting signals Failure to observe the specified maximum signal ratings can cause shock a fire hazard or damage to the devices c...

Page 17: ...at Start All Programs National Instruments NI DMM Documentation Figure 13 NI 4060 NI 4065 and NI 407x Signal Connections for Current Measurements Tip To measure current beyond the specifications of t...

Page 18: ...re Resistance Figure 14 NI 4050 Signal Connections for 2 Wire Resistance Measurements Figure 15 NI 4060 NI 4065 and NI 407x Signal Connections for 2 Wire Resistance Measurements 1 NI 4060 2 NI PXI PCI...

Page 19: ...ely by eliminating the effects of lead resistance Note The NI 4050 does not support 4 wire resistance measurements Figure 16 NI 4060 NI 4065 and NI 407x Signal Connections for 4 Wire Resistance Measur...

Page 20: ...ice under test and read the resulting voltage drop which is useful for testing diodes Figure 18 NI 4050 Signal Connections for Diode Measurements Figure 19 NI 4060 NI 4065 and NI 407x Signal Connectio...

Page 21: ...etween either of the inputs and the ground of the computer should never exceed 300 VDC or 300 Vrms The maximum current the NI 4065 can measure between the current inputs is 3 A DC or 3 Arms NI 4071 us...

Page 22: ...mation about using NI DMM in your applications NI DMM Examples The NI DMM examples are instructional tools that demonstrate some of the functionality of the DMMs you can use or integrate into your sys...

Page 23: ...t panel has five shrouded banana plug connectors and one 9 pin connector Refer to Figure 20 for the front panels for the NI 4060 NI 4065 and NI 407x Figure 20 Front Panel Connectors on the PXI PCI PCI...

Page 24: ...he digital signals on this connector should not operate beyond 0 5 to 5 5 V of your computer ground The trigger signals are TTL compatible PCMCIA Caution To measure current with the NI PCMCIA 4050 you...

Page 25: ...ots A hybrid slot can accept either PXI Express modules or PXI Express compatible modules but not standard PXI modules To determine if your chassis contains hybrid slots refer to your chassis document...

Page 26: ...tion that does not obstruct the fan mounting location s Add additional fans such as PCI fan cards and chassis fans to increase the air circulation inside the computer USB System Considerations The NI...

Page 27: ...at the angle shown in Figure 22 and locate the fuse holder Figure 22 Locating the Fuse on the NI PXI 4060 4 Insert a screwdriver into the slot on the fuse holder 5 Turn counterclockwise 6 Pull out the...

Page 28: ...NI PCI 4060 2 Power off the computer and remove the device 3 Remove the four screws that secure the top and bottom insulators onto the device as shown in Figure 23 Figure 23 Removing the Insulators f...

Page 29: ...PCI 4060 6 Remove the fuse and verify the fuse is blown by measuring discontinuity across the fuse using another meter 7 Press the new fuse into the silver holding fixture until you hear a snap Cautio...

Page 30: ...remove the device NI USB 4065 Disconnect the USB cable from the NI USB 4065 2 Locate the fuse holder on the NI 4065 NI PXI 4065 View the device at the angle shown in Figure 25 NI PCI PCIe 4065 View t...

Page 31: ...nal Instruments Corporation 31 NI Digital Multimeters Getting Started Guide Figure 26 Locating the Fuse on the NI PCI PCIe 4065 Figure 27 Locating the Fuse on the NI USB 4065 1 Fuse Holder 1 Fuse Hold...

Page 32: ...clockwise until it snaps shut NI 407x To replace the fuse complete the following steps NI PXI 4070 4071 4072 1 Remove all front panel connections from the NI PXI 407x 2 Power off the chassis and remov...

Page 33: ...ry the fuse loose from the slot 8 Verify the fuse is blown by measuring discontinuity across the fuse using another meter 9 Insert a new fuse into the fuse slot as shown in Figure 30 Figure 30 Replaci...

Page 34: ...l out the fuse holder and remove the 5 20 mm fuse 7 Verify the fuse is blown by measuring discontinuity across the fuse using another meter 8 Insert a new fuse into the holder and slide the holder bac...

Page 35: ...orth Mopac Expressway Austin Texas 78759 3504 National Instruments also has offices located around the world to help address your support needs For telephone support in the United States create your s...

Page 36: ...National Instruments trademarks Other product and company names mentioned herein are trademarks or trade names of their respective companies For patents covering National Instruments products refer t...

Page 37: ...Initialize Configure Hardware Basic Configuration Advanced Configuration Read Initiate Fetch Close Session NI DMM Programming Flow NI DMM...

Reviews: