SCXI-1503 Calibration Procedure
Test Equipment
National Instruments recommends using the equipment in Table 1 for adjusting and verifying the
SCXI-1503.
Test Conditions
The following setup and environmental conditions are required to ensure the SCXI-1503
meets
published specifications.
•
Keep connections to the module as short as possible. Long cables and wires act as antennae, which
can pick up extra noise that affects measurements.
•
Verify that all connections to the calibration system are secure.
•
Maintain an ambient temperature of 25 ±10 °C. The SCXI-1503 module temperature will be
greater than the ambient temperature. If you plan to operate the SCXI-1503 at a specific
temperature outside this range, calibrate the module at that temperature.
•
Keep relative humidity below 80%.
Table 1.
Recommended Equipment
Equipment
Recommended
Model
Parameter
Measured
Minimum Requirements
Calibrator
Fluke 5500A*
Analog Input
If this instrument is
unavailable, use a
calibrator that is accurate to
10 ppm (5 digits).
DMM
PXI-4071
Excitation Current
Analog Input
If this instrument is
unavailable, use a
multiranging 7½ digit
DMM.
SCXI Chassis
SCXI-1000
—
—
DAQ Device
NI PCI(e)/PXI(e) E, M
or X Series Device
—
16-bit DAQ device
that is supported by
NI-DAQmx 8.1 or later.
Terminal Block Connection
Accessory
SCXI-1306
—
—
Rear Connection Accessory
SCXI-1349
—
—
Terminal Accessory, 50-pin
CB-50
—
—
Ribbon Cable, 50-Pin
NB1
—
—
Cable between SCXI-1349
and E Series DAQ devices
SH68-68-EP
—
—
Cable between SCXI-1349
and M Series or X Series
DAQ devices
SHC68-68-EPM
—
—
* If using the Fluke 5500A calibrator, you must have the Fluke Fl5xmcal driver installed on the test computer.