SCXI-1503 Calibration Procedure
Initial Setup
Refer to the DAQ Getting Started Guides and the
SCXI Quick Start guide
for information about how to
install the software and hardware, and how to configure the module in MAX. Figure 1 shows a typical
SCXI setup for calibration using an NI PCI(e) DAQ device.
Figure 1.
Typical Setup for Verification and Adjustment Using an NI PCI(e) DAQ Device
1
SCXI-1306 Terminal Block
2
SCXI-1503 Module
3
SCXI-1000 Chassis
4
SCXI-1349 Adapter
5
SH68-68 EP Cable (for E Series DAQ Device)
SHC68-68 EPM Cable (for M or X Series DAQ Device)
6
Test Computer, with NI PCI(e) DAQ Device
7
DMM
8
Calibrator
9
Calibrator to SCXI-1306 Wiring
10 CB-50 to DMM Wiring
11 CB-50 Terminal Accessory
12 NB1 Ribbon Cable