1
2
3
4
1. Reference Antenna
2. Calibration Laser
3. Reference Antenna Fixture
4. Positioner
2.
Connect the opposite end of the 2.4 mm (m)-to-2.4 mm (m) cable connected to the
interior chamber rear bulkhead 1A port to the reference antenna port.
Note
The second 2.4 mm (m)-to-2.4 mm (m) cable connected to the interior
chamber rear bulkhead 2A port is utilized during system calibration. Use this
cable if your DUT has two input ports. Otherwise, leave the cable connected to
the 2A port until the system calibration procedure is complete.
3.
Close the chamber door.
4.
Set the
Elevation
in the mmWave OTA Test Positioner SFP to rotate the positioner so the
reference antenna points straight up toward the measurement antenna, as shown in the
previous figure.
Related Information
on page 18
mmWave OTA Test Positioner SFP
Attaching the Measurement Antenna
1.
Mount the measurement antenna to the measurement antenna fixture as shown in the
following figure.
mmWave OTA Validation Test Reference Architecture
|
© National Instruments
|