Installing the Module
Caution
Refer to the
Read Me First: Safety and Radio-Frequency Interference
document
before removing equipment covers or connecting/disconnecting any signal wires
A blue label stripe identifies the SCC-AI
XX
as an analog input module.
The label also displays the icon shown in Figure 1-1.
Figure 1-1.
SCC-AI
XX
Icon
You can plug the SCC-AI
XX
into any analog input socket on the SC-2345.
The socket you choose determines which E Series DAQ device channels
receive the SCC-AI
XX
signals.
For single-stage input conditioning, plug the SCC-AI
XX
into any socket
J(
X
+
1), where
X
is 0 to 7, and connect the input signals to the module as
described in the
section.
If you use the SCC-AI
XX
in a dual-stage configuration, the SCC-AI
XX
must be the first-stage module. Plug it into any socket J(
X
+
9) and plug the
second-stage SCC into socket J(
X
+
1), where
X
is 0 to 7. Connect the input
signals to the SCC-AI
XX
as described in the
section. The SC-2345 connects the output signals of the first-stage SCC to
the inputs of the second-stage SCC. An example of dual-stage conditioning
is an SCC-AI02 followed by an SCC-FV01 frequency input module.