NI 4070/4072 Calibration Procedure
28
ni.com
Figure 3.
Cable Connections for 2-Wire Resistance
30. Wait two minutes for the thermal EMF to stabilize.
31. Call
niDMM_ConfigureOffsetCompOhms
with
OffsetCompOhms
set to
NIDMM_VAL_OFFSET_COMP_OHMS_OFF
.
32. Output 10 k
Ω
on the calibrator with 2-wire compensation and external
sense turned on.
Table 10.
Fluke 5440 Cable Connections
Banana Plug Connector
(NI 4070/4072)
Banana Plug Color
(Fluke 5440 Cable)
Binding Post Label
(Fluke 5700A/5720A Calibrator)
HI
Red
OUTPUT HI
LO
Black
OUTPUT LO
(No connection)
Blue
V GUARD
HI
Red
OUTPUT HI SENSE
LO
Black
OUTPUT LO SENSE
(No connection)
Blue
V GUARD
HI
LO
AUX
I/O
3
00V
MAX
5V
MAX
AMPS
NI PXI-407x
6
1
/
2
-Digit FlexDMM
500V MA X
CAT I
HI
LO
1kV
MAX
3
A, 250V
MAX
INPUT
1 kV
MAX
5700A/5720A HI / LO SENSE
5440 C
ab
le
s
5700A/5720A HI / LO
5440 C
ab
le
s