National Instruments FieldPoint cFP-RTD-122 Operating Instructions Manual Download Page 9

© 

National Instruments Corp.

9

FP-RTD-122 and cFP-RTD-122

effects of the leads. The only residual errors are those caused by 
mismatching the EX+ and COM leads. Most RTDs have lead 
resistances within 5% of each other, so the compensation of the 
[c]FP-RTD-122 corrects for 95% or more of the errors introduced 
by lead resistances. This is a more accurate method than the typical 
bridge completion methods described in many reference books. 
The bridge methods not only have the same sensitivity to lead 
resistance mismatch, but also are effective only for temperatures 
very near those at which the bridge is balanced (usually 0 

°

C).

The temperature measurement accuracy specifications for the 
[c]FP-RTD-122 at the end of these instructions include the effects 
of a typical application using 10 m of 22 gauge copper wire 
(approximately 0.5 

 per lead), with 5% mismatch in the lead 

resistances. If you are using leads with greater resistances, the 
additional errors are approximately 3 

°

C per 

 of mismatch in the 

lead resistances for 100 

 RTDs, and 0.3 

°

C per 

 of mismatch in 

the lead resistance for 1,000 

 RTDs. For example, for 2 

 leads 

matched to 5% of each other, the lead resistance mismatch is 
5%

×

 = 0.1 

, which would cause 0.3 

°

C of error in 

measurements of a 100 

 RTD.

If you are using the [c]FP-RTD-122 with two-wire RTDs, the 
errors due to lead resistances are much greater because three-wire 
compensation is not used. With two-wire RTDs, the additional 
errors are approximately 3 

°

C per 

 of the sum of the lead 

resistances for 100

 RTDs, and 0.3 

°

C per 

 of the sum of lead 

resistances for 1,000

 RTDs. For example, a 1,000 

 two-wire 

RTD with 2 

 leads has a total lead resistance of 4 

 (2 

 per lead), 

which causes 1.2 

°

C of error.

Status Indicators

The [c]FP-RTD-122 has two green status LEDs, 

POWER

 and 

READY

. After you insert the [c]FP-RTD-122 into a terminal base 

or backplane and apply power to the connected network module, 
the green 

POWER

 indicator lights and the [c]FP-RTD-122 

informs the network module of its presence. When the network 
module recognizes the [c]FP-RTD-122, it sends initial 
configuration information to the [c]FP-RTD-122. After the 
[c]FP-AI-111 receives this initial information, the green 

READY

 

indicator lights and the module is in normal operating mode.

Summary of Contents for FieldPoint cFP-RTD-122

Page 1: ...how to install and use the National Instruments FP RTD 122 and cFP RTD 122 three wire RTD and resistance input modules referred to inclusively as the c FP RTD 122 For details on configuring and acces...

Page 2: ...r position 1 used for the FP RTD 122 module 2 Align the FP RTD 122 alignment slots with the guide rails on the terminal base 3 Press firmly to seat the FP RTD 122 on the terminal base When the module...

Page 3: ...he cFP RTD 122 prevent backward insertion 2 Press firmly to seat the cFP RTD 122 on the backplane 3 Using a number 2 Phillips screwdriver with a shank of at least 64 mm 2 5 in length tighten the capti...

Page 4: ...ield to the COM terminal Do not connect the shield to any of the wires at the signal end Taking RTD Measurements with the c FP RTD 122 The c FP RTD 122 has eight input channels All eight channels shar...

Page 5: ...xcitation wire to the EX terminal of the module and connect the other two wires to the SENSE and COM terminals Figure 4 Three Wire RTD Connections on One Channel Taking Measurements from Four Wire RTD...

Page 6: ...re resistance in ohms In this way you can take measurements from RTDs of types that the c FP RTD 122 does not directly support such as 120 nickel RTDs and from resistive devices other than RTDs You ca...

Page 7: ...RTD types are specified by material composition nominal resistance at 0 C and TCR The c FP RTD 122 can directly measure the temperature of platinum RTDs of either 100 or 1 000 nominal resistance These...

Page 8: ...arization curves measure the resistance with the c FP RTD 122 and convert the resistance to temperature in the manner suggested by the RTD vendor Three Wire Compensation of Lead Resistance Errors The...

Page 9: ...00 RTDs and 0 3 C per of mismatch in the lead resistance for 1 000 RTDs For example for 2 leads matched to 5 of each other the lead resistance mismatch is 5 2 0 1 which would cause 0 3 C of error in m...

Page 10: ...d the inter module communication bus There is no isolation between channels unless otherwise noted If any of the channels on a module are wired at a hazardous potential make sure that all other device...

Page 11: ...ble gases or fumes Specifications These specifications are typical for the range 40 to 70 C unless otherwise noted Gain error is calculated as a percentage of input signal value Input Characteristics...

Page 12: ...hannel is updated every 1 08 s Physical Characteristics Indicators Green POWER and READY indicators Weight FP RTD 122 140 g 4 8 oz cFP RTD 122 110 g 3 7 oz Power Requirements Power from network module...

Page 13: ...2 27 50 g 3 ms half sine 18 shocks at 6 orientations 30 g 11 ms half sine 18 shocks at 6 orientations Safety This product is designed to meet the requirements of the following standards of safety for...

Page 14: ...ref nsf Mechanical Dimensions Figure 7 shows the mechanical dimensions of the FP RTD 122 installed on a terminal base If you are using the cFP RTD 122 refer to the Compact FieldPoint controller user m...

Page 15: ...4 8521 Canada Toronto 905 785 0085 Canada Vancouver 514 685 7530 China 86 21 6555 7838 Czech Republic 420 2 2423 5774 Denmark 45 45 76 26 00 Finland 385 0 9 725 725 11 France 33 0 1 48 14 24 24 German...

Reviews: