BNC-2142 Installation Guide
4
www.natinst.com
Installing Your BNC-2142
Perform the following steps to connect your BNC-2142 to your DSA
device. Consult your computer user manual or technical reference manual
for specific instructions and warnings.
Note
You can either place the BNC-2142 on a workbench near the host computer or use
the available DIN rail-mount kit.
1.
Connect the BNC-2142 to your DSA device.
2.
Connect the field signals to the BNC connectors. Refer to the
Connecting Signals to Your BNC-2142 section earlier in this guide for
more information.
When you have finished using your BNC-2142, turn off any powered
external signals connected to your BNC-2142 before you turn off your
computer.
Figure 1.
Connecting the BNC-2142 to Your DSA Device
Warning
The BNC-2142 is not designed for input voltages greater than ±42.4 V, even if a
user-installed voltage divider reduces the voltage to within the input range of the DSA
device. Input voltages greater than ±42.4 V can damage the BNC-2142, any device
connected to it, and the host computer. Overvoltage can also cause an electric shock hazard
for the operator. National Instruments is not liable for damage or injury resulting from such
misuse.
1
BNC-2142
2
Shielded Cable
3
DSA Device
4
Personal Computer
3
4
1
2