Set the calibrator to Standby mode (STBY).

h)

Repeat steps d through g for each point in the array.

i)

Close the calibration session.

j)

Clear the task.

4.

Disconnect the NI 9252 from the calibrator.

EEPROM Update

When an adjustment procedure is completed, the NI 9252 internal calibration memory
(EEPROM) is immediately updated.

If you do not want to perform an adjustment, you can update the calibration date and onboard
calibration temperature without making any adjustments by initializing an external calibration,
setting the C Series calibration temperature, and closing the external calibration.

Reverification

Repeat the 

Verification

 section to determine the As-Left status of the device.

Note

 

If any test fails Reverification after performing an adjustment, verify that you

have met the Test Conditions before returning your device to NI. Refer to the

Worldwide Support and Services

 section for assistance in returning the device to NI.

Accuracy under Calibration Conditions

The values in the following table are based on calibrated scaling coefficients, which are stored
in the onboard EEPROM.

The following accuracy table is valid for calibration under the following conditions:

Ambient temperature 23 °C ± 5 °C

NI 9252 installed in slot 8 of an NI cDAQ-9178 chassis

Slots 1 through 7 of the NI cDAQ-9178 chassis are empty

Table 6. NI 9252 Accuracy under Calibration Conditions

Device

Percent of Reading (Gain Error)

Percentage of Range (Offset Error)

2

 

 

 

NI 9252

0.14%

0.025%

Note

 

For operational specifications, refer to the most recent 

NI 9252 Datasheet

online at 

ni.com/manuals

.

2

Range equals 10.5 V.

6

  | 

ni.com

  | 

NI 9252 Calibration Procedure

Summary of Contents for 9252

Page 1: ...6 Accuracy under Calibration Conditions 6 Revision History 7 Worldwide Support and Services 7 Software Calibrating the NI 9252 requires the installation of NI DAQmx 19 1 or later on the calibration system You can download NI DAQmx from ni com downloads NI DAQmx supports LabVIEW LabWindows CVI ANSI C and NET When you install NI DAQmx you only need to install support for the application software tha...

Page 2: ...ort for Visual Studio Reference information for NI DAQmx NET methods and NI DAQmx NET properties key concepts and a C enum to NET enum mapping table Test Equipment The following table lists the equipment recommended for the performance verification and adjustment procedures If the recommended equipment is not available select a substitute using the requirements listed Table 1 Recommended Equipment...

Page 3: ...ssis to the system safety ground a Attach a ring lug to a 14 AWG 1 6 mm wire b Connect the ring lug to the ground terminal on the side of the NI cDAQ 9178 using the ground screw c Attach the other end of the wire to the system safety ground 4 Install the module in slot 8 of the NI cDAQ 9178 chassis Leave slots 1 through 7 of the NI cDAQ 9178 chassis empty 5 Connect the NI cDAQ 9178 chassis to your...

Page 4: ...average samples a Create and configure an AI voltage channel according to the following NI 9252 AI Voltage Task table Table 2 NI 9252 AI Voltage Task Physical Channels Range Terminal Configuration cDAQMod8 AI0 7 10 5 V Differential b Configure the AI voltage channel timing according to the following NI 9252 Task Timing table Table 3 NI 9252 Task Timing Samples Per Channel Acquisition Mode Rate kS ...

Page 5: ... to the calibrator as shown in the Accuracy Verification section 3 Adjust the NI 9252 a Initialize a calibration session on the NI 9252 The default password is NI b Input the external temperature in degrees Celsius c Call the NI 9252 get C Series adjustment points function to obtain an array of recommended calibration voltages for the NI 9252 d Set the calibrator to a reference value determined by...

Page 6: ...ification after performing an adjustment verify that you have met the Test Conditions before returning your device to NI Refer to the Worldwide Support and Services section for assistance in returning the device to NI Accuracy under Calibration Conditions The values in the following table are based on calibrated scaling coefficients which are stored in the onboard EEPROM The following accuracy tab...

Page 7: ... Engineers Visit ni com services for information about the services NI offers Visit ni com register to register your NI product Product registration facilitates technical support and ensures that you receive important information updates from NI NI corporate headquarters is located at 11500 North Mopac Expressway Austin Texas 78759 3504 NI also has offices located around the world For support in t...

Page 8: ...s and third party legal notices in the readme file for your NI product Refer to the Export Compliance Information at ni com legal export compliance for the NI global trade compliance policy and how to obtain relevant HTS codes ECCNs and other import export data NI MAKES NO EXPRESS OR IMPLIED WARRANTIES AS TO THE ACCURACY OF THE INFORMATION CONTAINED HEREIN AND SHALL NOT BE LIABLE FOR ANY ERRORS U ...

Reviews: