30
|
ni.com
|
NI 9232 Operating Instructions and Specifications
Gain matching (calibrated)
Phase matching (maximum)
Passband frequency .......................... 0.4 ·
f
s
Flatness (peak-to-peak)
Frequency Band
20 Hz to 40.96 kHz
Typical
Maximum
channel-to-channel
25 mdB
120 mdB
Frequency Band
20 Hz to 40.96 kHz
channel-to-channel
(0.022°/kHz ·
f
in
)
+ 0.045°
module-to-module
(0.022°/kHz ·
f
in
)
+ 0.045° + (360° ·
f
in
/
f
M
)
Frequency Band
20 Hz to 20.48 kHz
20 Hz to 40.96 kHz
typical
70 mdB
75 mdB
maximum
75 mdB
*
100 mdB
†
*
For conformal coated version, the maximum flatness for 20 Hz to 20.48 kHz is 76 mdB.
†