NI 9219 Calibration Procedure
Current Accuracy Adjustment
Complete the following steps to adjust the current accuracy of the NI 9219.
1.
Connect the NI 9219 to the DMM and source as shown in Figure 4.
2.
Open a calibration session to your device using the DAQmx Initialize External Calibration VI. The
default password is
NI
.
3.
Place all channels on the NI 9219 in current mode using the NI 9219 instance of the DAQmx Setup
C Series Calibration VI.
LabVIEW Block Diagram
NI-DAQmx Function Call
Call
DAQmxInitExtCal
with the following
parameters:
deviceName
: The name of the device you want to
calibrate. This name can be found under Devices
and Interfaces in Measurements & Automation
Explorer.
password
:
NI
calHandle
:
&calHandle
LabVIEW Block Diagram
NI-DAQmx Function Call
Call
DAQmxSetup9219Cal
with the following
parameters:
rangeMin
:
–0.025
rangeMax
:
0.025
calHandle
:
calHandle
channelName
:
cDAQ1Mod8/a
i0:3
measType
:
DAQmx_Val_Current
bridgeConfig
:
DAQmx_Val_NoBridge