National Instruments Corporation
13
NI TB-2627 Installation Instructions
Caution
When hazardous voltages (>42.4 V
pk
/60 VDC) are present on any relay
terminal, safety low-voltage (<42.4 V
pk
/60 VDC) cannot be connected to any other
relay terminal.
Cold-Junction Sensor
Accuracy
1
From 0 °C to 15 °C ......................... ±1.0 °C
From 15 °C to 35 °C ....................... ±0.5 °C
From 35 °C to 55 °C ....................... ±1.0 °C
Output
From 0 °C to 55 °C ......................... 0.2 V to 0.024 V
1
Includes the combined effects of the temperature sensor accuracy and the temperature difference between the temperature
sensor and any screw terminal. The temperature sensor accuracy includes tolerances in all component values, effects caused
by temperature, loading, self-heating, and temperature gradients.