![National Instruments 776572-1600 User Manual Download Page 32](http://html1.mh-extra.com/html/national-instruments/776572-1600/776572-1600_user-manual_3513220032.webp)
Chapter 3
Configuring and Testing
©
National Instruments Corporation
3-7
6.
Select the channel(s) you want to verify. You can select a block of
channels by holding down the <Shift> key or multiple channels by
holding down the <Ctrl> key. Click
OK
.
7.
Enter the appropriate information on the
Settings
tab.
8.
Click the
Device
tab.
9.
Enter the appropriate information on the
Device
tab.
10. Click the
Test
button.
11. Click the
Start
button.
12. After you have completed verifying the channels, click the
Stop
button.
You have now verified the SCXI-1125 configuration and signal connection.
Note
For more information about how to further configure the SCXI-1125, or how to use
LabVIEW to configure the module and take measurements, refer to Chapter 4,
Verifying the Signal in Traditional NI-DAQ (Legacy)
This section discusses how to verify the signal in Traditional NI-DAQ
(Legacy) using channel strings and virtual channels.
Verifying the Signal Using Channel Strings
The format of the channel string is as follows:
ob
x
!
sc
y
!
md
z
!
channel
where
•
ob
x
is the onboard E Series DAQ device channel, with
x
representing
a particular channel where the multiplexed channels are sent. This
value is 0 for E Series DAQ device channel 0 in a single-chassis
system. In a multichassis or remote chassis system, the E Series DAQ
device channel
x
corresponds to chassis number
n
– 1, where E Series
DAQ device channel
x
is used for scanning the
n
th chassis in the
system.
•
sc
y
is the SCXI chassis ID, where
y
is the number you chose when
configuring the chassis.
•
md
z
is the slot position where the module is located, with
z
being the
particular slot number. The slots in a chassis are numbered from left to
right, starting with 1.
•
channel
is the channel that is sampled from module
z
.