background image

6703/6704 Calibration Procedure

4

www.natinst.com

Installing NI-DAQ

Since 6703/6704 boards are PC-based measurement devices, you must 
have the proper board driver installed in the calibration system before 
attempting any calibration. For the 6703/6704 calibration procedure, you 
will need the latest version of the National Instruments NI-DAQ driver 
installed on the calibration system. This driver, which configures and 
controls the DAQ device, is available at 

www.natinst.com

.

The driver supports a number of programming languages, including 
LabVIEW, LabWindows/CVI, Microsoft Visual C ++, Microsoft Visual 
Basic and Borland C ++. When you install the driver, you need only install 
support for the programming language that you intend to use.

For documentation that explains the use of the driver, install the following 
PDF files:

NI-DAQ Function Reference Manual for PC Compatibles

NI-DAQ User Manual for PC Compatibles

These two manuals provide detailed information on using the NI-DAQ 
driver. The function reference manual includes information on the 
functions in the driver. The user manual provides instructions on installing 
and configuring National Instruments DAQ devices and detailed 
information on creating applications that use the NI-DAQ driver. These 
manuals will be your primary references for writing your calibration utility. 
For further information on the products you will be calibrating, you may 
also want to install the device user manual.

You will also need a copy of the 

ni6704Cal.dll

ni6704Cal.lib

, and 

ni6704Cal.h

 files. The 

.dll

 provides calibration functionality that does 

not reside in the standard NI-DAQ driver, including the ability to protect the 
calibration constants, update the calibration date, and write to the factory 
calibration area. The factory calibration area and the calibration date should 
only be modified by a metrology laboratory or another facility that 
maintains traceable standards. You can access the functions in this 

.dll

 

through any 32-bit compiler. These files are available on the National 
Instruments Web site at 

www.natinst.com

.

Summary of Contents for 6703

Page 1: ...it This section also explains internal and external calibration Equipment and Other Test Requirements This section describes what you need to do before you can calibrate your device including an over...

Page 2: ...of your application indicate how often the 6703 6704 products need to be calibrated to maintain accuracy National Instruments recommends that you perform a complete calibration at least once every ye...

Page 3: ...the calibration and verification process to approximately 10 minutes However manual calibration and verification can take as long as 1 hour You can automate the adjustment and verification procedure i...

Page 4: ...e detailed information on using the NI DAQ driver The function reference manual includes information on the functions in the driver The user manual provides instructions on installing and configuring...

Page 5: ...edure The calibration procedure in the Calibration Process section later in this document provides step by step instructions on calling the appropriate calibration functions These calibration function...

Page 6: ...as by picking up extra noise that would affect calibration Maintain the temperature between 18 28 C Keep relative humidity below 80 Use shielded copper wire for all cable connections to the device It...

Page 7: ...his procedure can be easily expanded to verify all the voltage channels by performing the procedure once for each voltage channel To do this perform the procedure once for each value of the channel pa...

Page 8: ...he procedure once for each value of the channel parameter from 16 to 31 You will also need to connect the appropriate channel to your multimeter each time you repeat the procedure See the PCI PXI 6703...

Page 9: ...make offset and gain adjustments You perform offset adjustment first followed by gain adjustment You will then perform both offset and gain adjustment again to ensure proper accuracy This second set...

Page 10: ...the voltage on voltage channel 0 Decreasing the voltage value will increase the magnitude of the voltage on voltage channel 0 You have now calibrated the voltage output Calibrating Current Output 6704...

Page 11: ...on current channel 16 Decreasing the voltage value will increase the magnitude of the current on current channel 16 You have now calibrated the current output Saving Calibration Values Once you comple...

Page 12: ...ng tables are the specifications you should use when calibrating and verifying the 6703 6704 The tables show the specifications for 1 year and 24 hour calibration intervals Using the Tables The follow...

Page 13: ...r Limit V Lower Limit V Upper Limit V 0 0 0 0003125 0 0003125 0 0003125 0 0003125 10 0 9 9992675 10 0007325 9 9990975 10 0009025 Table 2 Current Output Specifications 6704 Only Test Point Value mA 24...

Page 14: ...r additional information on the software structure Voltage Output Verification Figure 1 Voltage Output Verification Look at the 6703 6704 specification table to determine the appropriate ranges for th...

Page 15: ...tput Verification 6704 Only Look at the 6703 6704 specification table to determine the appropriate ranges for the voltages and currents to be verified Get_Cal_Date deviceNumber Connect the multimeter...

Page 16: ...IBRATION_ENABLE ND_YES AO_VWrite deviceNumber 0 0 0 AO_VWrite deviceNumber 0 10 0 6703 AO_VWrite deviceNumber 16 voltage Adjust voltage until the DMM reads 10 0 V 6704 AO_VWrite deviceNumber 32 voltag...

Page 17: ...6704 Only AO_VWrite deviceNumber 16 0 1 Connect the multimeter to current channel 16 AO_VWrite deviceNumber 33 voltage Adjust voltage until the DMM reads 0 1 mA AO_VWrite deviceNumber 35 voltage Adju...

Page 18: ...libration Procedure 18 www natinst com Saving Calibration Values Figure 5 Saving Calibration Values Set_DAQ_Device_Info deviceNumber ND_CALIBRATION_ENABLE ND_NO AO_Calibrate deviceNumber 2 1 Copy_Cal...

Reviews: